Skip to the content

FDA Medical Device Recalls (openFDA)

Alere San Diego, Inc.: The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) Test Cassette, Model Number 12-807. Product Usage: Immunoassay for the determination of C-Reactive Protein (CRP). The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum.

Recall number
Z-1178-2013
Recalling firm
Alere San Diego, Inc.
Product
The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) Test Cassette, Model Number 12-807. Product Usage: Immunoassay for the determination of C-Reactive Protein (CRP). The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum.
Status
Terminated
Initiated
2013-03-06
Posted
2013-04-25
Terminated
2014-03-12
Reason
The Alere Cholestech LDX hsCRP cassette (PN 12-807) may have increased imprecision relative to the performance data in the package insert, which may cause differences in quantitative results for hsCRP compared to a reference method or repeat testing on the same patient sample.
Root cause
Nonconforming Material/Component
Action
Alere sent an Urgent Medical Device Recall letter dated March 6, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to block all stock of the affected lot in their warehouses; discontinue use and discard all stock on hand, complete the provided Customer Verification form to document disposal, contact their primary consignees and advise them of the required discontinuation and disposal of the affected lot. For question call 877-308-8289.
Quantity
2333 kits
Distribution
USA Nationwide Distribution
Lot, serial or model codes
Model Number 12-807. ¿ Lot/Unit Codes: 274632, 281847 ¿ Expiration date, or Expected shelf life: 5/5/13, 6/30/13
510(k) numbers
["K040579"]
PMA numbers
not on file
Product code
DCK
Firm FEI
1000125596
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register