FDA Medical Device Recalls (openFDA)
Alere San Diego, Inc.: The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) Test Cassette, Model Number 12-807. Product Usage: Immunoassay for the determination of C-Reactive Protein (CRP). The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum.
- Recall number
- Z-1178-2013
- Recalling firm
- Alere San Diego, Inc.
- Product
- The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) Test Cassette, Model Number 12-807. Product Usage: Immunoassay for the determination of C-Reactive Protein (CRP). The Alere Cholestech LDX high sensitivity C-Reactive Protein (hsCRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum.
- Status
- Terminated
- Initiated
- 2013-03-06
- Posted
- 2013-04-25
- Terminated
- 2014-03-12
- Reason
- The Alere Cholestech LDX hsCRP cassette (PN 12-807) may have increased imprecision relative to the performance data in the package insert, which may cause differences in quantitative results for hsCRP compared to a reference method or repeat testing on the same patient sample.
- Root cause
- Nonconforming Material/Component
- Action
- Alere sent an Urgent Medical Device Recall letter dated March 6, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to block all stock of the affected lot in their warehouses; discontinue use and discard all stock on hand, complete the provided Customer Verification form to document disposal, contact their primary consignees and advise them of the required discontinuation and disposal of the affected lot. For question call 877-308-8289.
- Quantity
- 2333 kits
- Distribution
- USA Nationwide Distribution
- Lot, serial or model codes
- Model Number 12-807. ¿ Lot/Unit Codes: 274632, 281847 ¿ Expiration date, or Expected shelf life: 5/5/13, 6/30/13
- 510(k) numbers
- ["K040579"]
- PMA numbers
- not on file
- Product code
- DCK
- Firm FEI
- 1000125596
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28