PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228

Recall number
Z-1177-2024
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228
Status
Open, Classified
Initiated
2024-02-06
Posted
2024-02-22
Terminated
not on file
Reason
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
Root cause
Under Investigation by firm
Action
Philips notified U.S customers by Urgent Medical Device Correction letter on 06-Feb-2024 via certified mailing via United States Postal Service. " Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk, and action to take: Circulate this Letter to all clinical staff and any service personnel that may service the system so that they are aware of the issue. " Keep this Urgent Medical Device Correction Letter with the documentation of the system until Philips implements the correction in your system. " If the Certeray generator fails, please stop using the system and contact the Customer Care Solutions Center. " Place a copy of this Letter in a visible place in the Control and Technical rooms (e.g., posted on the door). " Please complete and return the attached response form (on page 3) to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction and understanding of the issue and required actions to be taken. 5. Actions planned by Philips IGTS to correct the problem As a remedy, Philips will replace the Power Inverter in the Azurion generator cabinet in all affected Azurion systems through the implementation of Field Change Order (FCO72200544). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Quantity
27 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, State of Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom.
Lot, serial or model codes
OUS (1) Model: 722227 SN: UDI: 75 (01)00884838099227(21)75 122 (01)00884838099227(21)122 131 (01)00884838099227(21)131 132 (01)00884838099227(21)132 165 (01)00884838099227(21)165 116 (01)00884838099227(21)116 136 (01)00884838099227(21)136.' (2) Model: 722228 Serial Number UDI 168 (01)00884838099234(21)168 197 (01)00884838099234(21)197 154 (01)00884838099234(21)154 193 (01)00884838099234(21)193 165 (01)00884838099234(21)165 163 (01)00884838099234(21)163 139 (01)00884838099234(21)139 202 (01)00884838099234(21)202 141 (01)00884838099234(21)141 200 (01)00884838099234(21)200 174 (01)00884838099234(21)174 178 (01)00884838099234(21)178 147 (01)00884838099234(21)147 208 (01)00884838099234(21)208 171 (01)00884838099234(21)171 167 (01)00884838099234(21)167 194 (01)00884838099234(21)194 176 (01)00884838099234(21)176 228 (01)00884838086777(10)2.0.0.10015 161 (01)00884838099234(21)161
510(k) numbers
["K200917"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register