Skip to the content

FDA Medical Device Recalls (openFDA)

Pelton & Crane Company: Pelton & Crane Spirit 3000 dental operative unit with the TLC SAFEdrive option, 1727 Fuehauf Dr. Charlotte, NC 28273 Product Usage: The intended use of the dental operative unit is for supplying power and to serve as a base for other dental devices, such as a dental hand piece, air/water syringe, oral cavity evacuator, a suction operative unit, and other dental devices. The TLC SAFEdrive option is an optional feature that monitors the operation of the high speed hand piece and gives an audible and visual signal to let the user know if the hand piece is potentially damaged or heating up.

Recall number
Z-1177-2012
Recalling firm
Pelton & Crane Company
Product
Pelton & Crane Spirit 3000 dental operative unit with the TLC SAFEdrive option, 1727 Fuehauf Dr. Charlotte, NC 28273 Product Usage: The intended use of the dental operative unit is for supplying power and to serve as a base for other dental devices, such as a dental hand piece, air/water syringe, oral cavity evacuator, a suction operative unit, and other dental devices. The TLC SAFEdrive option is an optional feature that monitors the operation of the high speed hand piece and gives an audible and visual signal to let the user know if the hand piece is potentially damaged or heating up.
Status
Terminated
Initiated
2012-02-20
Posted
2012-03-06
Terminated
2012-03-28
Reason
Modification of the Pelton & Crane Spirit 3000 unit disabled the TLC SAFEdrive feature rendering it ineffective in detecting whether the handpiece is potentially damaged or heating up.
Root cause
Employee error
Action
Pelton & Crane sent an Urgent Medical Device Recall letter dated February 20, 2012 to all affected customers. The letter identifies the affected product, problem and actions to be taken. The letter states that a Pelton & Crane representative will be contacting the customers to schedule a service appointment with the end user of the device so that the affected TLC SAFEdrive circuit board can be replaced and returned to Pelton & Crane. Customers were instructed to complete the attached Recall Acknowledgement/Return Form and fax it to Pelton & Crane, attention Frank Ray, at (704) 587-7250. The letter also requested customers to mail the original Recall Acknowledgement/Return Form in the enclosed self addressed stamped envelope.
Quantity
2 units
Distribution
(USA) Nationwide Distribution
Lot, serial or model codes
Model #SET30, Catalog #3120, Serial numbers 4096 and 4099.
510(k) numbers
["K946187"]
PMA numbers
not on file
Product code
EIA
Firm FEI
1017522
Firm city
Charlotte
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register