FDA Medical Device Recalls (openFDA)
Canon Medical System, USA, INC.: Alphenix INFX-8000C, interventional fluoroscopic x-ray system
- Recall number
- Z-1176-2026
- Recalling firm
- Canon Medical System, USA, INC.
- Product
- Alphenix INFX-8000C, interventional fluoroscopic x-ray system
- Status
- Open, Classified
- Initiated
- 2025-12-19
- Posted
- 2026-01-27
- Terminated
- not on file
- Reason
- It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- Root cause
- Process design
- Action
- Canon Medical Systems USA issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2025 via letter. The notice explained the issue, risk to health, and requested the following: "Request to customers: Should the suspended device make an abnormal noise, or a sensor error is displayed before the corrective action is implemented on your system, please stop using the device and call your local Canon Service Team. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at intouchdispatch@us.medical.canon."
- Quantity
- 172 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Dominican Republic.
- Lot, serial or model codes
- serial numbers: W1C0632284, W4A1282279, W4A12Z2324, A2A2222381, A2A2392756, A3517001, A2A2352683, A2A2072087, A2A2072077, C1C2422647, W2A09Y2061, W2A10X2168, W4A11Y2144, A2A2232406, A2A2032038, W1D07Z2458, W2A09X2053, B2B1812645, A1C19Z2182, W4B1332362, B2A1622324, W4C1352385, W4A1292284, A1C19X2152, A1C1942058, A1C1972104, B2A1592233, B2A1792598, W2A1012081, W2A1052121, W4C1342374, B2B1872766, A4562105, W2A0972027, A1C2052199, A2A21X2339, W2A1012082, W4C1342369, W1D07Z2449, A1C1942065, A2A2132228, A2A2172285, A2A2117001, W1D06Z2357, A2A2282477, A2A2232405, W1B08X2034, A2A2182313, W4A1232210, A2A2262452, B2A16X2449, B2B1862743, B2B1872751, W1D0697014, W1C05X2229, B2A1542162, A2A21Z2364, A2A2392751, B2A1722527, A2A22Z2566, B4582115, W4C1372400, A2A2362690, A2A2252436, A2A2292510, B2A1582209, B2A1472014, W1D0732376, A2A2252439, W1D0732374, A1C1972105, A1C1972112, B4602126, W1B08Z2063, A2A2152265, A2A2132230, A2A2282491, A2A2262445, A2A22Y2550, A2A22X2528, W4A1212180, W4A1212176, W4A1212177, A1C1992141, W4A1182110, W4A11Z2163, W1D0812466, B2B18Y2816, B2B1832693, A1C1952078, A2A2062070, W2A09Z2077, A1C19Y2174, W1D0722372, A2A22X2518, A3542008, B2A1472020, W4A1242224, W4A1312348, A2A2172280, W4D1422466, W4A1142069, A2A2172287, A2A2342657, A2A2432872, B2A1552176, B2B1812657, B2B1872765, W4A1342373, B2A1682407, W4A11Z2170, W4C1342364, W4C1342365, W4A1172092, A2A2172281, W2A1072143, W4A1292281, A2A2242426, A2A2392761, A3612061, A1A2162240, A2A23Z2814, B2B1872760, A2A2262456, A1C2042197, B2A14Z2085, W4C1352377, W4A1252245, A1C19Z2178, A2A2292499, W4A1242221, W1D0872524, A1C19Z2188, A1C19X2158, A2A2052053, B2B1862731, B2A1752550, B2B1832678, B2B1852708, B2B1852710, B2B1852707, W1C0582207, W1C05Y2242, W1D0722370, W4A1272267, W2A1062127, B2A1742539, B2B1922842, A2A2062066, B2A16Y2469, W4A1122048, B2A1662369, W4A1122039, B5512160, A2A2122205, B2B1822664, B2B1822670, A2A2292498, W4D13Y2442, A2A2122197, A2A2192330, A2A2132216, A2A22X2522, A2A2432885, A2A2422855, W2A1012091, W1C05X2224, B2A1622310, A1C1972111, B2B1892787, A1C1952080, A1B18Y7002.
- 510(k) numbers
- ["K210900"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28