Skip to the content

FDA Medical Device Recalls (openFDA)

Canon Medical System, USA, INC.: Alphenix INFX-8000C, interventional fluoroscopic x-ray system

Recall number
Z-1176-2026
Recalling firm
Canon Medical System, USA, INC.
Product
Alphenix INFX-8000C, interventional fluoroscopic x-ray system
Status
Open, Classified
Initiated
2025-12-19
Posted
2026-01-27
Terminated
not on file
Reason
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Root cause
Process design
Action
Canon Medical Systems USA issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2025 via letter. The notice explained the issue, risk to health, and requested the following: "Request to customers: Should the suspended device make an abnormal noise, or a sensor error is displayed before the corrective action is implemented on your system, please stop using the device and call your local Canon Service Team. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at intouchdispatch@us.medical.canon."
Quantity
172 units
Distribution
Worldwide distribution - US Nationwide and the country of Dominican Republic.
Lot, serial or model codes
serial numbers: W1C0632284, W4A1282279, W4A12Z2324, A2A2222381, A2A2392756, A3517001, A2A2352683, A2A2072087, A2A2072077, C1C2422647, W2A09Y2061, W2A10X2168, W4A11Y2144, A2A2232406, A2A2032038, W1D07Z2458, W2A09X2053, B2B1812645, A1C19Z2182, W4B1332362, B2A1622324, W4C1352385, W4A1292284, A1C19X2152, A1C1942058, A1C1972104, B2A1592233, B2A1792598, W2A1012081, W2A1052121, W4C1342374, B2B1872766, A4562105, W2A0972027, A1C2052199, A2A21X2339, W2A1012082, W4C1342369, W1D07Z2449, A1C1942065, A2A2132228, A2A2172285, A2A2117001, W1D06Z2357, A2A2282477, A2A2232405, W1B08X2034, A2A2182313, W4A1232210, A2A2262452, B2A16X2449, B2B1862743, B2B1872751, W1D0697014, W1C05X2229, B2A1542162, A2A21Z2364, A2A2392751, B2A1722527, A2A22Z2566, B4582115, W4C1372400, A2A2362690, A2A2252436, A2A2292510, B2A1582209, B2A1472014, W1D0732376, A2A2252439, W1D0732374, A1C1972105, A1C1972112, B4602126, W1B08Z2063, A2A2152265, A2A2132230, A2A2282491, A2A2262445, A2A22Y2550, A2A22X2528, W4A1212180, W4A1212176, W4A1212177, A1C1992141, W4A1182110, W4A11Z2163, W1D0812466, B2B18Y2816, B2B1832693, A1C1952078, A2A2062070, W2A09Z2077, A1C19Y2174, W1D0722372, A2A22X2518, A3542008, B2A1472020, W4A1242224, W4A1312348, A2A2172280, W4D1422466, W4A1142069, A2A2172287, A2A2342657, A2A2432872, B2A1552176, B2B1812657, B2B1872765, W4A1342373, B2A1682407, W4A11Z2170, W4C1342364, W4C1342365, W4A1172092, A2A2172281, W2A1072143, W4A1292281, A2A2242426, A2A2392761, A3612061, A1A2162240, A2A23Z2814, B2B1872760, A2A2262456, A1C2042197, B2A14Z2085, W4C1352377, W4A1252245, A1C19Z2178, A2A2292499, W4A1242221, W1D0872524, A1C19Z2188, A1C19X2158, A2A2052053, B2B1862731, B2A1752550, B2B1832678, B2B1852708, B2B1852710, B2B1852707, W1C0582207, W1C05Y2242, W1D0722370, W4A1272267, W2A1062127, B2A1742539, B2B1922842, A2A2062066, B2A16Y2469, W4A1122048, B2A1662369, W4A1122039, B5512160, A2A2122205, B2B1822664, B2B1822670, A2A2292498, W4D13Y2442, A2A2122197, A2A2192330, A2A2132216, A2A22X2522, A2A2432885, A2A2422855, W2A1012091, W1C05X2224, B2A1622310, A1C1972111, B2B1892787, A1C1952080, A1B18Y7002.
510(k) numbers
["K210900"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register