Skip to the content

FDA Medical Device Recalls (openFDA)

Abiomed, Inc.: Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)

Recall number
Z-1175-2024
Recalling firm
Abiomed, Inc.
Product
Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
Status
Terminated
Initiated
2021-03-01
Posted
2024-02-22
Terminated
2025-03-06
Reason
Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
Root cause
Software design
Action
This is a retrospective report for the March 2021 Technical Bulletin, which described a Product Update; specifically the AIC Version 8.5 Software Update. The software update began rollout on March 4, 2021. According to the firm, Version 8.5 of the software contained bug fixes for the issues seen in version 8.4. The technical bulletin was posted to HeartRecovery.com on March 5, 2021.
Quantity
4,463 (US)
Distribution
US Nationwide distribution.
Lot, serial or model codes
Product Code 0042-0000-US UDI-DI 00813502010022; Product Code 0042-0010-US UDI-DI 00813502010985; Product Code 0042-0040-US UDI-DI 00813502011401; Software Versions V8.4 and V8.4.1
510(k) numbers
not on file
PMA numbers
["P140003"]
Product code
OZD
Firm FEI
1220648
Firm city
Danvers
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register