Skip to the content

FDA Medical Device Recalls (openFDA)

Canon Medical System, USA, INC.: System INFX-8000C

Recall number
Z-1175-2021
Recalling firm
Canon Medical System, USA, INC.
Product
System INFX-8000C
Status
Open, Classified
Initiated
2021-02-05
Posted
not on file
Terminated
not on file
Reason
The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.
Root cause
Software design
Action
A customer notification letter was sent to affected customers, which includes instructions to (1) manually operate the diaphragm blades of the collimator to adjust the X-ray irradiation field to the appropriate irradiation field position, (2) when adjusting the diaphragm blades, you can find this problem by checking the position of the diaphragm blades with respect to the image receiving surface displayed on the screen, (3) contact their Canon Service representative if they find any abnormalities regarding this issue in the operation of the Infinix System and if they have any questions. And includes a statement that Canon Medical Solutions USA will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Service, the details of which will be included in a subsequent communication to you or Canon field engineer site visit.
Quantity
272 in total
Distribution
Nationwide
Lot, serial or model codes
Model: CAS-880A, CAS-830B
510(k) numbers
["K152697"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register