Skip to the content

FDA Medical Device Recalls (openFDA)

ConMed Corporation: MINI INFANT < 3KG, RADIOTRANSLUCENT ELECTRODE, Philips Plug Style Connector, REF/Catalog Number 2602H Indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single pediatric use, only. The PadPro 2602 electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (non-invasive) cardiac pacemaker and the pediatric patients skin. The electrode is intended for use on pediatric infants whose weight is less than 10 kg. (22 lbs.).

Recall number
Z-1175-2018
Recalling firm
ConMed Corporation
Product
MINI INFANT < 3KG, RADIOTRANSLUCENT ELECTRODE, Philips Plug Style Connector, REF/Catalog Number 2602H Indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single pediatric use, only. The PadPro 2602 electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (non-invasive) cardiac pacemaker and the pediatric patients skin. The electrode is intended for use on pediatric infants whose weight is less than 10 kg. (22 lbs.).
Status
Terminated
Initiated
2018-02-23
Posted
not on file
Terminated
2020-09-17
Reason
Wire sets of certain devices were assembled with adhesive on the contact wires, which may impede conductivity and could cause failure
Root cause
Process design
Action
The firm, ConMed,issued an "URGENT: DEVICE RECALL" notice dated 2/28/2018 and business reply forms to customers on 2/28/2018 via courier service. The notice described the product, problem and actions to be taken. The Customers were instructed to do the following: Step 1: Review your inventory for any of the devices with the lot codes listed on the notice. Customers are encouraged to contact all of those departments within your facility and any other facilities within your organization that may have received affected products. It is imperative that all end users of these devices receive this notice and respond immediately. Step 2a: If you HAVE inventory of any of the devices with the lot codes listed on the notice, complete the business reply form and return it with the devices to: CONMED Corporation 525 French Road Utica, NY 13502 USA Attn: Ed Kovac Step 2b: If you DO NOT HAVE any affected devices to return, please complete the business reply form, indicating you have no devices and return by one of the means listed below: 1. Email to: padpro2602@conmed.com 2. Fax to: Field Action Support Team at 315-624-3225. 3. US mail, using enclosed postage-paid envelope, to: CONMED, 525 French Road, Utica, NY 13502; Attn: Field Action Support Team If you have any questions or requests, please don't hesitate to contact the Field Action Support Team at 1-800-448-6506, fax to 315-624-3225, or email padpro2602@conmed.com.
Quantity
18,720 units in total
Distribution
Worldwide Distribution: US (Nationwide) and countries of: New Zealand, Saudi Arabia, Switzerland, United Arab Emirates Uruguay & Republic of Macedonia (The former Yugoslav Republic of Macedonia).
Lot, serial or model codes
Lot Numbers: Y01221601, Y03241702, Y10121501, Y020713-17, Y04071501, & Y12021501.
510(k) numbers
["K002280"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
3001451571
Firm city
Utica
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register