Toshiba American Medical Systems Inc: Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000 is intended for use with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and transmission/output or images to external devices.

Recall number
Z-1175-2016
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000 is intended for use with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and transmission/output or images to external devices.
Status
Terminated
Initiated
2016-01-14
Posted
2016-03-21
Terminated
2017-02-24
Reason
When a user performed radiography using the wireless FPD, a message window appeared on the monitor stating that imaging transmission was not completed with no radiographic image. It also indicated to select the "OK" button to re-acquire the image data or to select the'Cancel" button to cancel the re-acquisition. As instructed the user selected "OK" and the same message window appeared. This same operation was repeated several times with the identical result - no image. Finally, the user selected "Cancel" and the re-acquisition was terminated.
Root cause
Radiation Control for Health and Safety Act
Action
Toshiba's planned action: 1."Toshiba America Medical Systems, Inc. (TAMS) will advise contact customers to stop using the system if any abnormalities are found using the system. 2. TAMS will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that TAMS will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. TAMS will implement this CAP by June 30, 2016. Customers with questions were advised to call 301-796-5910. For questions regarding this recall call 714-730-5000.
Quantity
28
Distribution
Nationwide distribution to PA, UT, FL, NY, LA, WV, NJ, OH, WI and TX.
Lot, serial or model codes
Serial number : USB1522023
510(k) numbers
["K133553"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register