Varian Medical Systems, Inc. Oncology Systems: TrueBeam and True Beam STx versions 1.0 through 1.5, Model number H19; Varian Medical Systems, Palo, Alto, CA 94304. TrueBeam Radiotherapy Delivery System is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Recall number
Z-1175-2012
Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Product
TrueBeam and True Beam STx versions 1.0 through 1.5, Model number H19; Varian Medical Systems, Palo, Alto, CA 94304. TrueBeam Radiotherapy Delivery System is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Status
Terminated
Initiated
2012-01-30
Posted
2012-03-06
Terminated
2014-04-16
Reason
An anomaly has been identified with the respiratory gating software of TrueBeam. When importing breath-hold gating protocols, the gating thresholds can be reset to default values, rather than retaining the thresholds established during planning.
Root cause
Device Design
Action
Urgent Medical Device Correction/Urgent Field Safety Notices were sent on January 30, 2012 by certified mail to US and Canadian consignees. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** Varian Sales, Marketing and Service Organizations also were informed via letter the same day. The Field Safety Notice with customer distribution list was e-mailed to Varian International Customer Support Managers for translation (if necessary) to be sent to affected sites with return proof of notification to be sent to Varian. The notice informed users of the recommended actions. Questions were directed to the Varian Help Desk 1-888-827-4265 (US and Canada).
Quantity
182 units ***12-12-12 AMENDED TO: 200 units***
Distribution
Nationwide Distribution (USA) and the countries of Australia, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, India, Israel, Italy, Netherlands, New Zealand, Norway, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, Turkey and the United Kingdom.
Lot, serial or model codes
Model H19: Code numbers:  H191182,H191035,H191150,H191100,H191103,H191040,H191072,H191065, H191089,H191116,H191068,H191160,H191141,H191113,H191027,H191145, H191026,H191093,H191054,H191189,H191129,H191090,H191197,H191022,H191063,H191010,H191162,H191078,H191098,H191142,H191066,H191005,H191018,H191076,H191188,H191048,H191137,H191064,H191153,H191044,H191151,H191046,H191136,H191067,H191031,H191077,H191158,H191028,H191156,H191088,H191132,H191009,H191069,H191085,H191186,H191101,H191094,H191039,H191124,H191073,H191183,H191008,H191133,H191016,H191003,H191071,H191175,H191152,H191168,H191099,H191166,H191029,H191036,H191155,H191092,H191034,H191043,H191045,H191074,H191170,H191130,H191055,H191147,H191178,H191114,H191107,H191164,H191047,H191015,H191091,H191019,H191038,H191163,H191203,H191110,H191061, H191049,H191179,H191096,H191079,H191106,H191017,H191154,H191006,H191024,H191108,H191149,H191052,H191199,H191023,H191117,H191033,H191104,H191131,H191011,H191118,H191087,H191109,H191053,H191187,H191126,H191191,H191042,H191080,H191161,H191105,H191012,H191025, H191115,H191111,H191140,H191020,H191013,H191169,H191127,H191086,H191144,H191095,H191125,H191002,H191112,H191021,H191138,H191014,H191135,H191139,H191097,H191032,H191102,H191119,H191128,H191060,H191001,H191198,H191057,H191122,H191146,H191157,H191143,H191083,H191180,H191121,H191181,H191123,H191081,H191165,H191176,H191056,H191120,H191082,H191041,H191070,H191134,H191148,H191030,H191058,H191007,H191177,H191051,H191172,H191059 and H191037,  *** 12-12-12, AMENDED INFO ***:  H191167, H191171, H191184, H191185, H191192, H191194, H191202, H191205, H191207, H191208, H191211, H191214, H191217, H191219, H191220, H191222, H191223, H191224***
510(k) numbers
["K111106"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register