FDA Medical Device Recalls (openFDA)
DePuy Mitek, Inc., a Johnson & Johnson Co.: ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Recall number
- Z-1174-2013
- Recalling firm
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Product
- ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Status
- Terminated
- Initiated
- 2013-03-22
- Posted
- 2013-04-25
- Terminated
- 2014-02-24
- Reason
- The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
- Root cause
- Labeling Change Control
- Action
- On March 22, 2013, firm notified consignees of recall via letter. Consignees were informed of the following: The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six weeks after implantation. As Blue and Violet ORTHOCORD free strand sutures share the same IFU, IFU-108147 will be revised. Consignees were requested to confirm the receipt of the letter by March 29, 2013.
- Quantity
- 206540
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- ALL from Jan 1 2010 - Corrected IFU implementation date
- 510(k) numbers
- ["K043298"]
- PMA numbers
- not on file
- Product code
- NEW
- Firm FEI
- 1221934
- Firm city
- Raynham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28