FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: Synchron Phenobarbital (PHE) Reagent, P/N 469785, Classification Product Code: KLT. Subsequent Product Codes: DIP, KLS, LCD, LCQ, LCR, LFN. PHE reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Drug Calibrator 1, is intended for quantitative determination of Phenobarbital concentration in human serum or plasma.
- Recall number
- Z-1174-2012
- Recalling firm
- Beckman Coulter Inc.
- Product
- Synchron Phenobarbital (PHE) Reagent, P/N 469785, Classification Product Code: KLT. Subsequent Product Codes: DIP, KLS, LCD, LCQ, LCR, LFN. PHE reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Drug Calibrator 1, is intended for quantitative determination of Phenobarbital concentration in human serum or plasma.
- Status
- Terminated
- Initiated
- 2009-08-11
- Posted
- 2012-03-06
- Terminated
- 2013-12-05
- Reason
- The recall was initiated because Beckman Coulter has confirmed reports of variability in results with the lots of Phenobarbital (PHE) reagent and calibrator.
- Root cause
- Nonconforming Material/Component
- Action
- Beckman Coulter sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated August 26, 2009 to all customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were instructed to share the recall information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. Customers were asked to complete and return a response form. Contact Beckman Coulter Customer Support Center (Hotline) at (800) 854-3633 for questions regarding this recall.
- Quantity
- 31,559 units total (30,617 units in the US)
- Distribution
- Worldwide Distribution--USA (nationwide) and the country of Canada.
- Lot, serial or model codes
- M810083 and earlier
- 510(k) numbers
- ["K955644"]
- PMA numbers
- not on file
- Product code
- DKB
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28