Horiba Instruments, Inc dba Horiba Medical: Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Product Usage: The ABX PENTRA ML is a software data management system intended for use on ABX Hematology Analyzers.
- Recall number
- Z-1173-2014
- Recalling firm
- Horiba Instruments, Inc dba Horiba Medical
- Product
- Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Product Usage: The ABX PENTRA ML is a software data management system intended for use on ABX Hematology Analyzers.
- Status
- Terminated
- Initiated
- 2014-02-14
- Posted
- 2014-03-11
- Terminated
- 2014-08-22
- Reason
- HORIBA Medical initiated this recall of all software versions of ABX PENTRA ML (Model A11A06152) due to a software bug with the potential of transmitting incorrect results to the Laboratory Information System.
- Root cause
- Software design
- Action
- Horiba sent an "URGENT SAFETY NOTICE letter datedFebruary 13, 2014 via by mail to customers. The letter identified the affected product, issue, impact, actions to be taken. Customers were instructed to share this information with laboratory staff and retain this notification as part of their laboratory documentation. Customers were requested to complete and return the enclosed response form via one of the following methods: Fax: 949-453-0600 Email: jim.knowles@horiba.com Mail: HORIBA Medical Attention: Jim Knowles, Recall Coordinator 34 Bunsen Drive Irvine, CA 92618. For questions contact your local Horiba Medical representative.
- Quantity
- 15
- Distribution
- USA Nationwide Distribution
- Lot, serial or model codes
- Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Shelf Life: N/A
- 510(k) numbers
- ["K053308"]
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 3002698356
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28