Skip to the content

FDA Medical Device Recalls (openFDA)

Horiba Instruments, Inc dba Horiba Medical: Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Product Usage: The ABX PENTRA ML is a software data management system intended for use on ABX Hematology Analyzers.

Recall number
Z-1173-2014
Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Product
Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Product Usage: The ABX PENTRA ML is a software data management system intended for use on ABX Hematology Analyzers.
Status
Terminated
Initiated
2014-02-14
Posted
2014-03-11
Terminated
2014-08-22
Reason
HORIBA Medical initiated this recall of all software versions of ABX PENTRA ML (Model A11A06152) due to a software bug with the potential of transmitting incorrect results to the Laboratory Information System.
Root cause
Software design
Action
Horiba sent an "URGENT SAFETY NOTICE letter datedFebruary 13, 2014 via by mail to customers. The letter identified the affected product, issue, impact, actions to be taken. Customers were instructed to share this information with laboratory staff and retain this notification as part of their laboratory documentation. Customers were requested to complete and return the enclosed response form via one of the following methods: Fax: 949-453-0600 Email: jim.knowles@horiba.com Mail: HORIBA Medical Attention: Jim Knowles, Recall Coordinator 34 Bunsen Drive Irvine, CA 92618. For questions contact your local Horiba Medical representative.
Quantity
15
Distribution
USA Nationwide Distribution
Lot, serial or model codes
Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Shelf Life: N/A
510(k) numbers
["K053308"]
PMA numbers
not on file
Product code
JQP
Firm FEI
3002698356
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register