Skip to the content

FDA Medical Device Recalls (openFDA)

Barco N.V.: Barco MDNC-3421 DISPLAY, DIAGNOSTIC RADIOLOGY

Recall number
Z-1172-2021
Recalling firm
Barco N.V.
Product
Barco MDNC-3421 DISPLAY, DIAGNOSTIC RADIOLOGY
Status
Terminated
Initiated
2021-01-22
Posted
not on file
Terminated
2022-11-18
Reason
A number of MDNC-3421 displays were shipped while not being DICOM-compliant. These displays will not pass DICOM acceptance tests.
Root cause
Process change control
Action
On January 22, 2021 the firm notified the importer in the USA, Barco Inc.
Quantity
1 unit
Distribution
WI.
Lot, serial or model codes
Model Number: MDNC-3421, Serial number 2590263533, UDI/DI 05415334028388.
510(k) numbers
["K170837"]
PMA numbers
not on file
Product code
PGY
Firm FEI
3004455125
Firm city
Kortrijk
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register