Skip to the content

FDA Medical Device Recalls (openFDA)

Accumetrics Inc: VerifyNow IIb/IIIa Test, Catalog No. 85310, 10-Test Kit and Catalog No. 85011, 25-Test kit. Product Usage: The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow IIb/IIIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) IIb/IIIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow Ilb/Illa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician.

Recall number
Z-1172-2014
Recalling firm
Accumetrics Inc
Product
VerifyNow IIb/IIIa Test, Catalog No. 85310, 10-Test Kit and Catalog No. 85011, 25-Test kit. Product Usage: The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow IIb/IIIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) IIb/IIIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow Ilb/Illa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician.
Status
Terminated
Initiated
2014-02-13
Posted
2014-03-11
Terminated
2014-08-23
Reason
Accumetrics is recalling the VerifyNow IIb/IIIa test because it may result in the reporting of an erroneous low platelet aggregation unit (PAU) result. An erroneous low PAU result may cause a low percent inhibition calculation or low baseline PAU.
Root cause
Nonconforming Material/Component
Action
Accumetrics sent an Urgent Medical Device Recall letter dated February 13, 2014 to all affected customers. The recall letter informs the customers of the problems identified and the actions to be taken. Cusotomers are instructed to complete the attached Customer Account Tracking Form and return it within 10 days by fax, email or mail. Customers with questions are instructed call Accumetrics Customer Support at (800) 643-1640 option 2 or email at support@accumetrics.com.
Quantity
576 units total (524 units in US)
Distribution
USA Nationwide Distribution
Lot, serial or model codes
Lot/Serial No.  WC0182B, WC0182C, WC0182A, WC0182D, WC0182E, WC0182F, WC0182G, WC0180C, WC0180E, WC0180F, WC0181A, WC0181C, WC0181D, WC0181F, WC0180A, WC0180B, WC0180D, WC0181B, WC0181E, WC0181G, WC0181H, WC0181J, WC0181K.
510(k) numbers
["K011337","K013596","K992531"]
PMA numbers
not on file
Product code
JOZ
Firm FEI
3001376288
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28