FDA Medical Device Recalls (openFDA)
Accumetrics Inc: VerifyNow IIb/IIIa Test, Catalog No. 85310, 10-Test Kit and Catalog No. 85011, 25-Test kit. Product Usage: The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow IIb/IIIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) IIb/IIIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow Ilb/Illa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician.
- Recall number
- Z-1172-2014
- Recalling firm
- Accumetrics Inc
- Product
- VerifyNow IIb/IIIa Test, Catalog No. 85310, 10-Test Kit and Catalog No. 85011, 25-Test kit. Product Usage: The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow IIb/IIIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) IIb/IIIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow Ilb/Illa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician.
- Status
- Terminated
- Initiated
- 2014-02-13
- Posted
- 2014-03-11
- Terminated
- 2014-08-23
- Reason
- Accumetrics is recalling the VerifyNow IIb/IIIa test because it may result in the reporting of an erroneous low platelet aggregation unit (PAU) result. An erroneous low PAU result may cause a low percent inhibition calculation or low baseline PAU.
- Root cause
- Nonconforming Material/Component
- Action
- Accumetrics sent an Urgent Medical Device Recall letter dated February 13, 2014 to all affected customers. The recall letter informs the customers of the problems identified and the actions to be taken. Cusotomers are instructed to complete the attached Customer Account Tracking Form and return it within 10 days by fax, email or mail. Customers with questions are instructed call Accumetrics Customer Support at (800) 643-1640 option 2 or email at support@accumetrics.com.
- Quantity
- 576 units total (524 units in US)
- Distribution
- USA Nationwide Distribution
- Lot, serial or model codes
- Lot/Serial No. WC0182B, WC0182C, WC0182A, WC0182D, WC0182E, WC0182F, WC0182G, WC0180C, WC0180E, WC0180F, WC0181A, WC0181C, WC0181D, WC0181F, WC0180A, WC0180B, WC0180D, WC0181B, WC0181E, WC0181G, WC0181H, WC0181J, WC0181K.
- 510(k) numbers
- ["K011337","K013596","K992531"]
- PMA numbers
- not on file
- Product code
- JOZ
- Firm FEI
- 3001376288
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28