FUJIFILM Healthcare Americas Corporation: Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.

Recall number
Z-1171-2024
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.
Status
Open, Classified
Initiated
2024-01-03
Posted
2024-02-21
Terminated
not on file
Reason
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
Root cause
Software design
Action
On January 3, 2024, the firm notified customers via URGENT MEDICAL DEVICE CORRECTION letters. Fujifilm will offer to correct the formula that calculates the LV Mass (2D Bullet) in the Synapse database. Until the update is scheduled and completed, customers may continue to safely use the Pediatric Workflow modules listed in the notice as long as the LV Mass (2D Bullet) equation is not the main factor for a diagnosis. Customers should coordinate with the Fujifilm service engineer to schedule the correction of their device when contacted. Please contact Fujifilm's field action coordinator, Kotei Aoki, via phone at 765-246-2931 or email at kotei.aoki@fujifilm.com, if you have any further questions regarding this field action.
Quantity
13
Distribution
US domestic distribution nationwide. International distribution to Israel.
Lot, serial or model codes
UDI-DI 854904006008 Software versions 7.3.000, 7.2.200, 7.2.100, 7.2.000, 7.1.000, 7.0.000
510(k) numbers
["K190232"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000513161
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register