Skip to the content

FDA Medical Device Recalls (openFDA)

Hologic, Inc: Brand Name: Sertera Biopsy Device Product Name: Sertera Biopsy Device Model/Catalog Number: Sertera-14 Software Version: N/A Product Description: Instrument, Biopsy Component: N/A

Recall number
Z-1170-2025
Recalling firm
Hologic, Inc
Product
Brand Name: Sertera Biopsy Device Product Name: Sertera Biopsy Device Model/Catalog Number: Sertera-14 Software Version: N/A Product Description: Instrument, Biopsy Component: N/A
Status
Open, Classified
Initiated
2025-01-15
Posted
2025-02-13
Terminated
not on file
Reason
Devices presented a condition in which the inner needle separated from the hub and produced a longer throw of the inner needle or dislodgement.
Root cause
Process control
Action
On January 15, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Our goal is to create as little disruption to your practice as possible. Below outlines the next steps for customers with affected product(s): " Please quarantine any product(s) from the impacted lots you currently have in your inventory. Impacted lots can be returned through IQVIA as part of the voluntary recall. Please note that you will be issued a credit on your account for the affected product(s). " Refer to the Customer Response Form Instructions to identify your product inventory that is subject to this action. " Confirm receipt of this communication by completing the online Customer Response Form. Please complete and submit the online Customer Response Form even if you no longer have the affected product in inventory. Hologic has partnered with IQVIA to conduct follow up communications should no response be received to this letter. " Hologic will issue a Return Goods Authorization upon receipt of the Customer Response Form and the return of all affected products to Hologic. " If you wish to reorder product, please place a new order via your preferred procurement method. We value your business and thank you for your cooperation. If you have any questions about this communication, please contact BreastHealth.Support@hologic.com or 1-877-371-4372.
Quantity
55,335 units
Distribution
Domestic: Nationwide Distribution
Lot, serial or model codes
Model No- Sertera-14; UDI- 15420045504066; Lot Numbers: E23B23RL, E23B23RM, E23C07RD, E23C07RF, E23C31RC, E23D12RT, E23D12RU, E23D27RW, E23E03RJ, E23E12RT, E23E12RU, E23E29RL, E23E30RM, E23F02RG, E23F07RL, E23F07RM, E23F27R, E23F29RL, E23F29RM, E23F29RN, E23G12RK, E23G25R, E23G31RL, E23H02R, E23H02RA, E23H02RB, E23H28RA, E23H31RF, E23H31RG, E23K02RL, E23K10R, E23K10RA, E23K10RB, E23K31RU, E23L06RK, E23L10RB, E23L22RV, E23L28RB, E23L28RD, E23M21RV, E24A08RL, E24A08RM, E24A26RD, E24B01RF, E24B07RK, E24B22RC, E24B28R, E24B28RA, E24C07RA, E24C21R, E24D26RF, E24E02RD, E24E15RA, E24E22R, E24E22RA, E24E24RB.
510(k) numbers
["K150169"]
PMA numbers
not on file
Product code
KNW
Firm FEI
1000120743
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register