Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Captivator II Single-Use Polypectomy Snares Outer box UPN, Inner Pouch UPN: M00561191 (Box 10), M00561190; M00561221 (Box 10), M00561220; M00561222 (Box 20), M00561220; M00561223 (Box 40), M00561220; M00561233 (Box 40), M00561230.

Recall number
Z-1169-2021
Recalling firm
Boston Scientific Corporation
Product
Captivator II Single-Use Polypectomy Snares Outer box UPN, Inner Pouch UPN: M00561191 (Box 10), M00561190; M00561221 (Box 10), M00561220; M00561222 (Box 20), M00561220; M00561223 (Box 40), M00561220; M00561233 (Box 40), M00561230.
Status
Open, Classified
Initiated
2021-01-19
Posted
not on file
Terminated
not on file
Reason
Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
Root cause
Process change control
Action
On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
Quantity
335,788 units total (Domestic: 222,882, Foreign: 112,906)
Distribution
US nationwide distribution.
Lot, serial or model codes
Outer box GTIN: 08714729855934 Inner pouch GTIN: 08714729861263 Lot/Batch#: 25053639  Outer box GTIN: 08714729855903 Inner pouch GTIN: 08714729861294 Lot/Batch#: 24948407, 24670592, 24679619, 24676929, 24939628  Outer box GTIN: 08714729861300 Inner pouch GTIN: 087147298 61294 Lot/Batch#: 24770325, 24952552, 24665104  Outer box GTIN: 08714729861317 Inner pouch GTIN: 08714729861294 Lot/Batch#: 24679614, 25054403  Outer box GTIN: 08714729861348 Inner pouch GTIN: 08714729861324 Lot/Batch#: 24772995
510(k) numbers
["K133987"]
PMA numbers
not on file
Product code
FDI
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register