FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255

Recall number
Z-1169-2018
Recalling firm
Beckman Coulter Inc.
Product
AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255
Status
Completed
Initiated
2018-01-19
Posted
not on file
Terminated
not on file
Reason
Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.
Root cause
Device Design
Action
Beckman Coulter plans to replace the belt via Service Modification during the next planned service maintenance visit. At that time, the customer will be informed of the belt replacement by the service engineer.
Quantity
98 units in the US
Distribution
Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States
Lot, serial or model codes
All systems up to the following Serial Numbers: ODL25120 Revision T; SN3101, ODL25125 Revision W; SN1508, ODL25250 Revision T; SN2416, ODL25255 Revision W; SN4617
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register