FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255
- Recall number
- Z-1169-2018
- Recalling firm
- Beckman Coulter Inc.
- Product
- AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255
- Status
- Completed
- Initiated
- 2018-01-19
- Posted
- not on file
- Terminated
- not on file
- Reason
- Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.
- Root cause
- Device Design
- Action
- Beckman Coulter plans to replace the belt via Service Modification during the next planned service maintenance visit. At that time, the customer will be informed of the belt replacement by the service engineer.
- Quantity
- 98 units in the US
- Distribution
- Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States
- Lot, serial or model codes
- All systems up to the following Serial Numbers: ODL25120 Revision T; SN3101, ODL25125 Revision W; SN1508, ODL25250 Revision T; SN2416, ODL25255 Revision W; SN4617
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28