Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Captivator Single-Use Polypectomy Snares, Captiflex Single-Use Polypectomy Snares. Outer box UPN, Inner Pouch UPN: M00561311 (Box 10), M00561310 M00562301 (Box 10), M00562300; M00562321 (Box 10), M00562320; M00562341 (Box 10), M00562340; M00562451 (Box 10), M00562450; M00562401 (Box 10), M00562400; M00562402 (Box 40), M00562400; M00562422 (Box 40), M00562420; M00562471 (Box 10), M00562470.

Recall number
Z-1168-2021
Recalling firm
Boston Scientific Corporation
Product
Captivator Single-Use Polypectomy Snares, Captiflex Single-Use Polypectomy Snares. Outer box UPN, Inner Pouch UPN: M00561311 (Box 10), M00561310 M00562301 (Box 10), M00562300; M00562321 (Box 10), M00562320; M00562341 (Box 10), M00562340; M00562451 (Box 10), M00562450; M00562401 (Box 10), M00562400; M00562402 (Box 40), M00562400; M00562422 (Box 40), M00562420; M00562471 (Box 10), M00562470.
Status
Open, Classified
Initiated
2021-01-19
Posted
not on file
Terminated
not on file
Reason
Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
Root cause
Process change control
Action
On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
Quantity
335,788 units total (Domestic: 222,882, Foreign: 112,906)
Distribution
US nationwide distribution.
Lot, serial or model codes
Outer box GTIN: 08714729019251 Inner pouch GTIN: 08714729747680 Lot/Batch#: 24725319, 24782695,   Outer box GTIN: 08714729019312 Inner pouch GTIN: 08714729747833 Lot/Batch#: 24842597, 24812577,   Outer box GTIN: 08714729019336 Inner pouch GTIN: 08714729747857 Lot/Batch#: 24889757, 24719202, 24709536, 24773287,   Outer box GTIN: 08714729019350 Inner pouch GTIN: 08714729747871 Lot/Batch#: 24782693, 24732964, 24889755, 24780939,   Outer box GTIN: 08714729071068 Inner pouch GTIN: 08714729747970 Lot/Batch#: 24882357, 24782689, 24782687, 24729758,   Outer box GTIN: 08714729019411 Inner pouch GTIN:08714729747932 Lot/Batch#: 24961643  Outer box GTIN:08714729501640 Inner pouch GTIN: 08714729747932 Lot/Batch#: 24961645, 24989260, 24844193,   Outer box GTIN:08714729501657 Inner pouch GTIN:08714729747956 Lot/Batch#: 25010054,   Outer box GTIN:08714729019459 Inner pouch GTIN:08714729747994 Lot/Batch#: 25016793.
510(k) numbers
["K202478"]
PMA numbers
not on file
Product code
FDI
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register