Skip to the content

FDA Medical Device Recalls (openFDA)

Vance Products Incorporated: Cook Urological Black Silicone Filiform Double Pigtail Ureteral Stent Set, 7.0Fr/24cm, sterile, Order Number (REF) 133724.

Recall number
Z-1168-2007
Recalling firm
Vance Products Incorporated
Product
Cook Urological Black Silicone Filiform Double Pigtail Ureteral Stent Set, 7.0Fr/24cm, sterile, Order Number (REF) 133724.
Status
Terminated
Initiated
2007-06-29
Posted
not on file
Terminated
2008-06-11
Reason
The radiographic band may detach from the stent and remain in the patient after removal of the stent.
Root cause
Other
Action
Consignees were notified of the recall via letter dated 6//29/07, which instructed them to quarantine stock on hand and to fax number on hand back to Cook for return authorization.
Quantity
75
Distribution
Worldwide, including USA, Canada, and Japan.
Lot, serial or model codes
Lots U1629501, U1629502 and U1631782.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FAD
Firm FEI
1825146
Firm city
Spencer
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register