Beckman Coulter Inc.: AU5800 Clinical Chemistry Analyzer The Beckman Coulter AU5800 Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. Applications include colorimetric, turbidimetric, latex agglutination, homogeneous enzyme immunoassay, and in selective electrode.
- Recall number
- Z-1167-2013
- Recalling firm
- Beckman Coulter Inc.
- Product
- AU5800 Clinical Chemistry Analyzer The Beckman Coulter AU5800 Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. Applications include colorimetric, turbidimetric, latex agglutination, homogeneous enzyme immunoassay, and in selective electrode.
- Status
- Terminated
- Initiated
- 2013-03-20
- Posted
- 2013-04-24
- Terminated
- 2014-09-16
- Reason
- Beckman Coulter, Inc. initiated this recall because there is a software error that has the potential to cause the reporting of erroneous patient results because of misidentification by the AU5800 system.
- Root cause
- Software design
- Action
- Beckman Coulter sent an Urgent Product Correction letter during the week of March 18, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were informed of the software limitation and provides instructions to be taken (temporary workaround) until the software is upgraded and released. Customer contact information provided: "Call Center Hotline at 800 854-3633. New customers will receive a copy of the customer notification at the time of installation by the Field Service Personnel until revised software that corrects this issue is released.
- Quantity
- 490
- Distribution
- Worldwide Distribution - USA, including Canada and Puerto Rico and the states of NJ, TX, CA, UT, KS, MA, NY, MI, GA, VA, IN, WI, IL, OH, NV, NC and PA. and the countries of Australia, Belgium, Czech Republic, Japan, Germany, Brazil, China, Italy, Ireland, Israel, India, Hong Kong, Malaysia, Netherlands, Spain, Switzerland, France, Taiwan, Turkey, United Kingdom and Vietnam.
- Lot, serial or model codes
- All Instrument Serial Numbers with Software Versions 4.11 and Prior Model Numbers AU5811-01, AU5821-01, AU5831-01, AU5841-01, AU5811-02, AU5821-02, AU5831-02, AU5841-02, AU5811-03, AU5821-03, AU5831-03, AU5841-03, AU5811-04, AU5821-04, AU5831-04, AU5841-04, AU5811-06, AU5821-06, AU5831-06, AU5841-06 Part Numbers A94906, A94911, A94916, A91921 A94907, A94912, A94917, A94922 A94908, A94913, A94918, A94923 A94909, A94914, A91949, A94924 A94910, A94915, A94920, A94925 Model Numbers and Part Numbers listed above respecitively.
- 510(k) numbers
- ["K112412"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28