Philips And Neusoft Medical Systems Co., Ltd.: NeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computed X-ray, Tomography system. The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients

Recall number
Z-1165-2013
Recalling firm
Philips And Neusoft Medical Systems Co., Ltd.
Product
NeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computed X-ray, Tomography system. The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients
Status
Terminated
Initiated
2013-03-04
Posted
2013-04-23
Terminated
2013-11-22
Reason
User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.
Root cause
Use error
Action
The firm, Neusoft Medical Systems Co., Ltd., sent an "URGENT DEVICE CORRECTION" NeuViz Dual and NeuViz 16 letter, dated Mar. 4, 2013, to their US consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to review this information with all members of their staff and retain a copy of the letter with the equipment Instruction For Use. Customers with questions should contact the Service Support Department by email to nms-service@neusoft.com
Quantity
23 units
Distribution
Nationwide distribution: USA including states of: California, Connecticut, North Carolina, Nebraska, Ohio, South Carolina, Texas, and Puerto Rico.
Lot, serial or model codes
Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120020, N16E120022, N16E120023, N16E120024, N16E120052, N16E120053, and N16E130002.
510(k) numbers
["K090173","K092742"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3003575350
Firm city
Shenyang
Firm state
Liaoning
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register