Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration

Recall number
Z-1164-2013
Recalling firm
GE Healthcare, LLC
Product
GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration
Status
Terminated
Initiated
2011-10-06
Posted
2013-04-23
Terminated
2014-02-04
Reason
GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts in attenuation corrected images. This may lead to a potential for misdiagnosis.
Root cause
Software design
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated October 5, 2011, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees were instructed to contact their local GE Healthcare Service Representative if they have questions or concerns. They may also contact GE Healthcare Call Center , United States or Canada at 800-437-1171. Other countries 262-896-2890. For questions regarding this recall call 262-513-4122.
Quantity
11
Distribution
Worldwide Distribution - USA including GA, and NY; and countries of: Canada, Israel, Italy and Switzerland.
Lot, serial or model codes
Mfg Lot or Serial # 0000CZC7105FJM000 00000CZC827NRN000 00000CZC8283V4000 0000CZC8283V47000 0000CZC8283V72001 0000CZC8283TX5001 00000000211423001 0000CZC8161XJT000 0000CZC94154WV003 0000CZC83942HL000 0000CZC8396CC9001
510(k) numbers
["K051673"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register