Skip to the content

FDA Medical Device Recalls (openFDA)

Johnson & Johnson Surgical Vision Inc: TECNIS Toric 1-Piece IOL, Model Number: ZCT150 +23.5D SE 1.50D CYL - Product Usage: intended to be placed in the capsular bag.

Recall number
Z-1163-2021
Recalling firm
Johnson & Johnson Surgical Vision Inc
Product
TECNIS Toric 1-Piece IOL, Model Number: ZCT150 +23.5D SE 1.50D CYL - Product Usage: intended to be placed in the capsular bag.
Status
Terminated
Initiated
2021-01-27
Posted
not on file
Terminated
2022-11-09
Reason
Due to the release of nonconforming Intraocular Lenses (IOLs).
Root cause
Process change control
Action
On 01/27/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via confirmed mail to customers informing them that due to investigation results, following a customer complaint, which confirmed a hyperopic patient outcome post implantation for an Intraocular Lens (IOL). The investigation for this complaint determined that this occurred due to inadvertent release of nonconforming product (Product labeled as 23.5D but measured to be 17.5D). Customers are instructed to: 1) Compare their inventory against the attached list on page 3. The IOL serial number is displayed on the end of each individual unit carton (see page 3 for label example). 2) Discontinue using and remove from their inventory all affected IOLs. 3) Complete and return the attached Customer Reply Form to acknowledge receipt of this Recall Notice, even if they have no inventory. The Recalling Firm requires this information for reconciliation purposes with regulatory agencies. The Recalling Firm has pre-filled the Customer Reply Form with the IOL serial number(s) that was(were) shipped to them and marked accordingly where product has already been returned. The completed Customer Reply Form should be emailed to RegCompliOne@its.jnj.com or faxed to 1-714-460-0466 within 3 business days of receipt of the recall notification letter. 4) Should customers have product to return with the serial number listed on page 3, they are asked to contact Customer Support at 1-877-266-4543 to arrange pick up of affected product. 5) The Recall Notification Letter should be shared with anyone who needs to be aware within the customer's organization or to any organization where the potentially affected products have been transferred. Any questions contact the Public Contact/Director, Quality Assurance at 949-480-7526 or email jturpins@its.jnj.com
Quantity
27 lenses
Distribution
U.S. Nationwide distribution in the states of AL, AZ, CA, FL, GA, IL, LA, MI, MN, NC, NY, OH, PA, SC, TN, TX, UT and WA.
Lot, serial or model codes
Model Number: ZCT150 UDI/GTIN Codes: (01)05050474552210(17)221123(21)8144121847 5050474552210 Lot Number: 8144121847
510(k) numbers
not on file
PMA numbers
["P980040S039"]
Product code
MJP
Firm FEI
3003843509
Firm city
Santa Ana
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register