Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: MOSAIQ Oncology Information System

Recall number
Z-1162-2025
Recalling firm
Elekta, Inc.
Product
MOSAIQ Oncology Information System
Status
Open, Classified
Initiated
2025-01-15
Posted
2025-02-12
Terminated
not on file
Reason
Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.
Root cause
Software design
Action
On January 15, 2025 Elekta issued a recall notification to affected consignees via Email. Elekta asked consignees to take the following actions. 1. Customers included in the scope of the Field Safety Notice, should not calculate shifts from 2D verification images taken at a non-zero patient support angle (couch angle). 2. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 3. Advise the appropriate personnel, working with this product, on the content of this letter. 4. You are required to either acknowledge receipt of this notification via the Elekta Care" Community or complete this form and return it to Elekta immediately upon receipt, but no later than within 30 days.
Quantity
64 units
Distribution
US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela
Lot, serial or model codes
Software versions /UDI: 3.1.2.0 (01)07340201500026 3.1.3.0 (01)07340201500026; 3.1.3.1 (01)07340201500026; 3.1.3.2 (01)07340201500026; 3.1.3.3 (01)07340201500026; 3.1.3.4 (01)07340201500026; 3.2.1.0 (01)07340201500071; 3.2.1.1 (01)07340201500071; 3.2.1.2 (01)07340201500071; 3.2.1.3 (01)07340201500071; 3.2.1.4 (01)07340201500071
510(k) numbers
["K223229"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register