bioMerieux Inc: OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D

Recall number
Z-1161-06
Recalling firm
bioMerieux Inc
Product
OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D
Status
Terminated
Initiated
2006-03-28
Posted
2006-06-22
Terminated
2008-03-07
Reason
Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle ID's
Root cause
Other
Action
The recalling firm telephoned customers who ordered the MB bottles and use the OBSERVA software ver. 2.0 to inform them of the issues. Recall letters were issued to all customers via overnight mail on 5/22/06 informing them a field representative would be visiting to make a short-term correction to the system. The letter also informed the customer a permanent correction to the software was scheduled for release later in the year.
Quantity
51 units
Distribution
WorldwideDistribution--Nationwide to medical centers. Government distribution was made to MN and CA, and military distribution was made to OK. Foreign distribution was made to Thailand, The Netherlands, Germany, Greece, Portugal, and France.
Lot, serial or model codes
Software version R.02.00.17
510(k) numbers
not on file
PMA numbers
not on file
Product code
MDB
Firm FEI
1950204
Firm city
Hazelwood
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register