Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of insulin levels in human serum and plasma (EDTA) using the Access Immunoassay Systems. Beckman Coulter, Inc., Chaska, MN 55318.

Recall number
Z-1160-2008
Recalling firm
Beckman Coulter Inc
Product
Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of insulin levels in human serum and plasma (EDTA) using the Access Immunoassay Systems. Beckman Coulter, Inc., Chaska, MN 55318.
Status
Terminated
Initiated
2007-09-19
Posted
2008-06-25
Terminated
2012-05-09
Reason
False negative results when used to test serum samples (as compared to plasma samples).
Root cause
Other
Action
On September 19, 2007, Beckman Coulter mailed to its consignees an Urgent Product Corrective Action Letter informing them that the affected lots of the product were giving false negative results when used to test serum samples (as compared to plasma samples). The letter instructed consignees to discontinue using the affected lots of the product, and to re-test using an EDTA plasma sample if there was any discordance between the results of the serum sample (if it was obtained using an affected lot of the product) as compared with the clinical presentaion of the patient and the results of other diagnostic tests. The letter also instructed consignees to return, within 10 days, the Response Sheet .
Quantity
8,205 units in the US; 332 in Canada
Distribution
Worldwide: including USA and Canada.
Lot, serial or model codes
Part Number 33410, Lot Numbers: 714362 (Exp dt. 3/31/09), 716129 (Exp dt. 6/30/09), 718258 (Exp dt. 7/31/09), 519096 (Exp dt. 10/31/07), 521487 (Exp dt. 2/28/08), 616274 (Exp dt. 7/31/08), 617890 (Exp dt. 7/31/08), 617891 (Exp dt. 7/31/08) and 620204 (Exp dt. 12/31/08)
510(k) numbers
["K953084"]
PMA numbers
not on file
Product code
CFP
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register