FDA Medical Device Recalls (openFDA)

Medtronic Gastroenterology / Urology: Medtronic Zinetics Manometric Catheter, Model 9012P2271. Water perfused, motility catheter. Distributed by Medtronic, Inc. 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604 USA.

Recall number
Z-1159-06
Recalling firm
Medtronic Gastroenterology / Urology
Product
Medtronic Zinetics Manometric Catheter, Model 9012P2271. Water perfused, motility catheter. Distributed by Medtronic, Inc. 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604 USA.
Status
Terminated
Initiated
2006-04-06
Posted
2006-06-22
Terminated
2007-01-16
Reason
Mislabeling- The 8 ports of the manometric catheter were incorrectly labeled 1,3,5,7,8,2,4,6 from the distal to the proximal end of the catheter. The correct port configuration is sequentially 1 through 8, from the distal to the proximal end of the catheter.
Root cause
Other
Action
Recall initiated to International accounts on 04/05/06 and within US on 04/06/06 via an Urgent Recall Notification. The letter informs them of the manufacturing defect, requests return for credit of any unused inventory and requests accounting of all previously used catheters. Distributors are instructed to sent a letter to their customers to request product return.
Quantity
94 catheters (US - 4; OUS - 90)
Distribution
Worldwide Distribution-including states of IL, NC, NY. and countries of Austria, Chile, Finland, Germany, Greece, Hong Kong, Italy, Japan, Lebanon, Norway, Peru, Portugal, Saudi Arabi, Spain
Lot, serial or model codes
Lots 0001391 and 0001464
510(k) numbers
not on file
PMA numbers
not on file
Product code
KLA
Firm FEI
1000135098
Firm city
Shoreview
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register