FDA Medical Device Recalls (openFDA)
Cardinal Health 200, LLC: Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A
- Recall number
- Z-1158-2023
- Recalling firm
- Cardinal Health 200, LLC
- Product
- Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A
- Status
- Open, Classified
- Initiated
- 2022-11-18
- Posted
- 2023-02-22
- Terminated
- not on file
- Reason
- Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
- Root cause
- Nonconforming Material/Component
- Action
- Cardinal Health issued Urgent: Product Correction letter and Acknowledgement Form was sent overnight on November 18, 2022 to the direct consignees. Letter states reason for recall, health risk and action to take: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: gmb-FieldCorrectiveAction@cardinalhealth.com For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. We sincerely apologize for any inconvenience this notice may have caused you and your staff. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332
- Quantity
- 36 EA
- Distribution
- US Nationwide distribution in the states of GA, PA.
- Lot, serial or model codes
- UDI-DI: 50195594381965(cs) 10195594381967(ea) Lot Number: 706265 Exp. Date: 5/1/2024
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRS
- Firm FEI
- 3001236905
- Firm city
- Waukegan
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.