BioFire Diagnostics, LLC: BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 - Product Usage: intended for use with FilmArray systems.

Recall number
Z-1158-2021
Recalling firm
BioFire Diagnostics, LLC
Product
BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 - Product Usage: intended for use with FilmArray systems.
Status
Terminated
Initiated
2021-01-26
Posted
not on file
Terminated
2022-05-02
Reason
The firm has identified an increased risk of false positive Pseudomonas aeruginosa when BCID/BCID2 Panels are used with specific BD BACTEC" blood culture vials.
Root cause
Other
Action
On or around January 25, 2021, the firm send an UGENT: MEDICAL DEVICE CORRECTION letter via email to all affected consignees. The firm's recall strategy includes notification of all customers who have ordered the BCID or BCID2 Panel, within the last 12 months, using the aforementioned Customer Letters. The effectiveness of this communication will be verified by monitoring confirmation of the Acknowledgment of Receipt from each affected customer.
Quantity
5,220 kits
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Afghanistan, Angola, Anguilla , Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Brazil, Brunei. Daruss., Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Czech Republic, Dominican Rep, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Utd/Arab Emir/, Vietnam.
Lot, serial or model codes
All non-expired BCID and BCID2 Panel lots, if used with the BD BACTEC Culture Media vials Catalog No. 442023 and 442020.
510(k) numbers
["K193519"]
PMA numbers
not on file
Product code
PEN
Firm FEI
3002773840
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register