Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Leksell Gamma Knife Perfexion, Article #715000. Teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.

Recall number
Z-1158-2010
Recalling firm
Elekta, Inc.
Product
Leksell Gamma Knife Perfexion, Article #715000. Teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.
Status
Terminated
Initiated
2007-07-02
Posted
2010-03-17
Terminated
2010-03-25
Reason
Need to modify the closing speed of the shielding doors in the event of an emergency exit
Root cause
Other
Action
A field action (FCO TN-07-0007) has been performed to modify the shielding door speed on all sites in clinical use. A second Field Change Order 200 069 "Replacement of Door Servo Board" dated 9/16/08 was released.
Quantity
19
Distribution
Worldwide Distribution -- USA, including states of AR, AZ, CA, FL, IL, MI, MN, MS, NJ, OH, OR, PA, UT, VA and countries of France and United Kingdom.
Lot, serial or model codes
Serial Numbers: 6002, 6003, 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, and 6030.
510(k) numbers
["K063512"]
PMA numbers
not on file
Product code
IWB
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register