FDA Medical Device Recalls (openFDA)
Bio-Rad Laboratories: BIOPLEX 2200 EBV IgM KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1350 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is a multiplex flow immunoassay intended for the qualitative detection of two (2) analytes; Epstein-Barr Virus Viral Capsid Antigen (EBV VCA) IgM antibodies and Heterophile antibodies in human serum.
- Recall number
- Z-1158-2008
- Recalling firm
- Bio-Rad Laboratories
- Product
- BIOPLEX 2200 EBV IgM KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1350 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is a multiplex flow immunoassay intended for the qualitative detection of two (2) analytes; Epstein-Barr Virus Viral Capsid Antigen (EBV VCA) IgM antibodies and Heterophile antibodies in human serum.
- Status
- Terminated
- Initiated
- 2007-11-13
- Posted
- 2008-06-11
- Terminated
- 2011-05-23
- Reason
- False negative results due to reagent packs exhibiting low signal.
- Root cause
- Nonconforming Material/Component
- Action
- On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.
- Quantity
- 996 kits
- Distribution
- Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.
- Lot, serial or model codes
- Lot Numbers: 943712 Exp. 7/30/07 and 943713 Exp. 4/30/08
- 510(k) numbers
- ["K062213"]
- PMA numbers
- not on file
- Product code
- LSE
- Firm FEI
- 3008272466
- Firm city
- Redmond
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28