Skip to the content

FDA Medical Device Recalls (openFDA)

Orthopedic Alliance LLC: Orthopedic Alliance Spine System (alias "Blue & Gold") Part Numbers: BGC3010 Screw Cap BGR0040 Straight Rod 5.5 x 40mm BGR0050 Straight Rod 5.5 x 50mm BGR0060 Straight Rod 5.5 x 60mm BGR0070 Straight Rod 5.5 x 70mm BGR0080 Straight Rod 5.5 x 80mm BGR0090 Straight Rod 5.5 x 90mm BGR0100 Straight Rod 5.5 x 100mm BGR0110 Straight Rod 5.5 x 110mm BGR0120 Straight Rod 5.5 x 120mm BGR0130 Straight Rod 5.5 x 130mm BGR0140 Straight Rod 5.5 x 140mm BGR0150 Straight Rod 5.5 x 150mm BGR0200 Straight Rod 5.5 x 200mm BGR0300 Straight Rod 5.5 x 300mm BGR0400 Straight Rod 5.5 x 400mm BGR0500 Straight Rod 5.5 x 500mm BGS5535 5.5 x 35mm Polyaxial Screw BGS5540 5.5 x 40mm Polyaxial Screw BGS5545 5.5 x 45mm Polyaxial Screw BGS6035 6.0 x 35mm Polyaxial Screw BGS6040 6.0 x 40mm Polyaxial Screw BGS6045 6.0 x 45mm Polyaxial Screw BGS6050 6.0 x 50mm Polyaxial Screw BGS6535 6.5 x 35mm Polyaxial Screw BGS6540 6.5 x 40mm Polyaxial Screw BGS6545 6.5 x 45mm Polyaxial Screw BGS6550 6.5 x 50mm Polyaxial Screw BGS7035 7.0 x 35mm Polyaxial Screw BGS7040 7.0 x 40mm Polyaxial Screw BGS7045 7.0 x 45mm Polyaxial Screw BGS7050 7.0 x 50mm Polyaxial Screw BGS7055 7.0 x 55mm Polyaxial Screw BGS7535 7.5 x 35mm Polyaxial Screw BGS7540 7.5 x 40mm Polyaxial Screw BGS7545 7.5 x 45mm Polyaxial Screw A pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.

Recall number
Z-1156-2013
Recalling firm
Orthopedic Alliance LLC
Product
Orthopedic Alliance Spine System (alias "Blue & Gold") Part Numbers: BGC3010 Screw Cap BGR0040 Straight Rod 5.5 x 40mm BGR0050 Straight Rod 5.5 x 50mm BGR0060 Straight Rod 5.5 x 60mm BGR0070 Straight Rod 5.5 x 70mm BGR0080 Straight Rod 5.5 x 80mm BGR0090 Straight Rod 5.5 x 90mm BGR0100 Straight Rod 5.5 x 100mm BGR0110 Straight Rod 5.5 x 110mm BGR0120 Straight Rod 5.5 x 120mm BGR0130 Straight Rod 5.5 x 130mm BGR0140 Straight Rod 5.5 x 140mm BGR0150 Straight Rod 5.5 x 150mm BGR0200 Straight Rod 5.5 x 200mm BGR0300 Straight Rod 5.5 x 300mm BGR0400 Straight Rod 5.5 x 400mm BGR0500 Straight Rod 5.5 x 500mm BGS5535 5.5 x 35mm Polyaxial Screw BGS5540 5.5 x 40mm Polyaxial Screw BGS5545 5.5 x 45mm Polyaxial Screw BGS6035 6.0 x 35mm Polyaxial Screw BGS6040 6.0 x 40mm Polyaxial Screw BGS6045 6.0 x 45mm Polyaxial Screw BGS6050 6.0 x 50mm Polyaxial Screw BGS6535 6.5 x 35mm Polyaxial Screw BGS6540 6.5 x 40mm Polyaxial Screw BGS6545 6.5 x 45mm Polyaxial Screw BGS6550 6.5 x 50mm Polyaxial Screw BGS7035 7.0 x 35mm Polyaxial Screw BGS7040 7.0 x 40mm Polyaxial Screw BGS7045 7.0 x 45mm Polyaxial Screw BGS7050 7.0 x 50mm Polyaxial Screw BGS7055 7.0 x 55mm Polyaxial Screw BGS7535 7.5 x 35mm Polyaxial Screw BGS7540 7.5 x 40mm Polyaxial Screw BGS7545 7.5 x 45mm Polyaxial Screw A pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.
Status
Terminated
Initiated
2013-03-07
Posted
2013-04-22
Terminated
2014-03-26
Reason
Orthopedic Alliance is recalling Blue & Gold Implants and Instruments because the Blue & Gold system is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product peformance failures that could cause patient harm due to implant breakage, loosening, or inadequate sterilization.
Root cause
Device Design
Action
A recall letter dated March 7, 2013 was sent to one customer who purchased the Blue & Gold Implants and instruments. The letter informed the customer that the Blue & Gold system is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product performance failures that could cause patient harm due to implant breakage, loosening, or inadequate sterilization. Customer is informed of the actions to be taken. Customer is instructed to complete the customer response form and return it by FAX to (858) 764-9739.
Quantity
3,300 units
Distribution
Distributed in Texas.
Lot, serial or model codes
All lots are being recalled.
510(k) numbers
["K033826"]
PMA numbers
not on file
Product code
MNI
Firm FEI
3004022256
Firm city
Murrieta
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28