Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corp: CircuCool Fluid Circulation Pump used in the Chilli Cooled Ablation System, Model Number M00480050, manufactured by Barnant ThermoFisher Scientific, Barrington, IL. Distributed by Boston Scientific Corporation, San Jose, CA. The Chilli Cooled Ablation System is indicated for cardiac electrophysiological mapping, delivering diagnostic pacing stimuli and radiofrequency ablation of mappable ventricular tachycardias attributable to ischemic heart disease or cardiomyopathy in patients who have failed drug therapy.

Recall number
Z-1156-2010
Recalling firm
Boston Scientific Corp
Product
CircuCool Fluid Circulation Pump used in the Chilli Cooled Ablation System, Model Number M00480050, manufactured by Barnant ThermoFisher Scientific, Barrington, IL. Distributed by Boston Scientific Corporation, San Jose, CA. The Chilli Cooled Ablation System is indicated for cardiac electrophysiological mapping, delivering diagnostic pacing stimuli and radiofrequency ablation of mappable ventricular tachycardias attributable to ischemic heart disease or cardiomyopathy in patients who have failed drug therapy.
Status
Terminated
Initiated
2009-09-22
Posted
2010-03-17
Terminated
2010-12-27
Reason
Product failure-- Product may cease operation as a result of potential speed control board failure when the device is shut off, and may not turn back on. A component on the circuit board fails unexpectedly because of a power surge, causing a diode to burn out.
Root cause
Nonconforming Material/Component
Action
The firm has issued notification letters via Federal express with reply verification tracking forms. A follow up letter and telephone survey as needed will be sent to follow up with non-responders. The firm is asking its consignees to return any recalled product. If there are any questions, please call 1.408.935.4952.
Quantity
75 units
Distribution
US and to Spain, France, Germany, Italy, Russia, Poland, the Netherlands, Israel and Belgium
Lot, serial or model codes
Serial numbers:  A9003399 to A9003418, B08003046 to B08003050, B08004180 to B08004209, D08003578 to D08003597, E09002815 to E09002839, G08002514 to G08002533, and K08005118 to K08005137
510(k) numbers
not on file
PMA numbers
["P980003"]
Product code
LPB
Firm FEI
3001236349
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register