Skip to the content

FDA Medical Device Recalls (openFDA)

St. Jude Medical: NurseMate with Physio Module; Remote Review, Monitoring and Charting Station; USA EPMedSystems, Inc. Cooper Run Executive Park, 575 Route 73 North, Unit D, West Berlin, NJ 08090. Outer box labeled as Smiths Medical. The NurseMate Remote Review Station is an integrated review station for the EP WorkMate Computer recording system that allows a separate user to review and edit patient study date stored on the SP Work Mate System and monitor patient data from the EP WorkMate during a Patient study.

Recall number
Z-1156-2009
Recalling firm
St. Jude Medical
Product
NurseMate with Physio Module; Remote Review, Monitoring and Charting Station; USA EPMedSystems, Inc. Cooper Run Executive Park, 575 Route 73 North, Unit D, West Berlin, NJ 08090. Outer box labeled as Smiths Medical. The NurseMate Remote Review Station is an integrated review station for the EP WorkMate Computer recording system that allows a separate user to review and edit patient study date stored on the SP Work Mate System and monitor patient data from the EP WorkMate during a Patient study.
Status
Terminated
Initiated
2009-02-13
Posted
2009-04-10
Terminated
2009-05-04
Reason
Installation problem: there is a risk that the monitor may not be adequately secured to the stand, resulting in the monitor falling off which may cause injury.
Root cause
Nonconforming Material/Component
Action
St. Jude Medical received a recall notice dated January 16, 2009 from its supplier, Smiths Medical PM, Inc. St. Jude Medical prepared and sent their own recall letter dated February 13, 2009 to affected customers. The recall letter described the issue and possible associated risks and provided instructions to customers. A St. Jude Medical representative will contact customers to facilitate replacement of the fasteners. Questions should be directed to your St. Jude Medical sales representative.
Quantity
4 units
Distribution
Nationwide Distribution -- CA, NV, and FL.
Lot, serial or model codes
Smith Medical Part Number: 87-1785-0000; Serial Numbers: AM08080015, AM08080016, AM08080017, AM08080018, AM08080019, AM08100004, AM08100005, AM08100006, AM08100007 and AM08100008.
510(k) numbers
["K070706"]
PMA numbers
not on file
Product code
DQK
Firm FEI
2126673
Firm city
West Berlin
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register