Canon Medical System, USA, INC.: The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.
- Recall number
- Z-1155-2024
- Recalling firm
- Canon Medical System, USA, INC.
- Product
- The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.
- Status
- Open, Classified
- Initiated
- 2023-11-15
- Posted
- 2024-02-16
- Terminated
- not on file
- Reason
- VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
- Root cause
- Component change control
- Action
- On 11/15/23 correction notices were distributed to customers who were asked to do the following: 1) Should this issue occur please attempt to restart the system and contact your service representative. 2) A recalling firm representative will contact you to schedule a time to replace the flat panel detector on the console. The parts required to address the defect are expected to be available no earlier that December 2023. 3) Complete and return the customer reply form via email to RAFFAIRS@US.MEDICAL.CANON For questions, please contact our InTouch team at intouchdispatch@us.medical.canon or 800-421-1968, or 800.521.1968.
- Quantity
- 2
- Distribution
- US: CA, LA
- Lot, serial or model codes
- UDI-DI: 04987670100208. Serial Number: BGE2292016, BGE2362017
- 510(k) numbers
- ["K182415"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28