FDA Medical Device Recalls (openFDA)
Zimmer Manufacturing B.V.: Product 57 consists of all product code: JDI and same usage: Item no: 763401000 PRESSFIT FEM ST FX 10 X 115 763401100 PRESSFIT FEM ST FX 11 X 1 763401200 PRESSFIT FEM ST FX 12 X 1 763401300 PRESSFIT FEM ST FX 13 X 1 763401400 PRESSFIT FEM ST FX 14 X 1 763401500 PRESSFIT FEM ST FX 15 X 1 For use in total or hemi hip arthroplasty
- Recall number
- Z-1155-2016
- Recalling firm
- Zimmer Manufacturing B.V.
- Product
- Product 57 consists of all product code: JDI and same usage: Item no: 763401000 PRESSFIT FEM ST FX 10 X 115 763401100 PRESSFIT FEM ST FX 11 X 1 763401200 PRESSFIT FEM ST FX 12 X 1 763401300 PRESSFIT FEM ST FX 13 X 1 763401400 PRESSFIT FEM ST FX 14 X 1 763401500 PRESSFIT FEM ST FX 15 X 1 For use in total or hemi hip arthroplasty
- Status
- Terminated
- Initiated
- 2016-01-11
- Posted
- 2016-03-12
- Terminated
- 2017-04-13
- Reason
- As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
- Root cause
- Packaging
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
- Quantity
- 0
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- lot no.: 62012579 62086025 62086025N 62055993 62264285 62362923 62264285N 62151833 62284483 62362921 62284483N 62224932 62284484 62340084 61886896 62043474 62141793 62284485 62362930 62284485N 62240754 62293494
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JDI
- Firm FEI
- 1000132037
- Firm city
- Mercedita
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28