GE Healthcare Clinical Systems: GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201
- Recall number
- Z-1155-2008
- Recalling firm
- GE Healthcare Clinical Systems
- Product
- GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201
- Status
- Terminated
- Initiated
- 2007-09-04
- Posted
- 2008-06-12
- Terminated
- 2012-05-29
- Reason
- Inaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt by the user to activate the biopsy guidelines while this probe is in use will cause incorrect guidelines to be displayed on the image. The incorrect guidelines are recognized by their orientation that is left/right reversed from t
- Root cause
- Software change control
- Action
- GE sent consignees a "GE Urgent Medical Device Correction" letter dated 8/14/07. The letter informed the customers about the problem, offer a short-term workaround and informed them that a representative will schedule a visit to upgrade the software.
- Quantity
- 62
- Distribution
- Worldwide Distribution - USA (NY & AL), Korea, China, India, Brazil, Mexico, Germany, Greece, Italy, Poland, Russia, Saudi Arabia, Spain, and Czech Republic.
- Lot, serial or model codes
- Systems with software version 4.0.5, 4.1.1 and 4.1.2. Serial numbers: 26022WS4, 26023WS2, 26346WS7, 2641WS8, 31001WS1, 34381WS4, 36460WS4, 36461WS2, 43734WS3, 43735WS0, 43736WS8, 43737WS6, 43994WS3, 43995WS0, 43998WS4, 44000WS8, 44001WS6, 44002WS4, 44003WS2, 44004WS0, 44005WS7, 44011WS5, 44012WS3, 44013WS1, 44015WS6, 44016WS4, 44018WS0, 44019WS8, 44020WS6, 44100WS6, 44153WS5, 44157WS6, 44159WS2, 44270WS7, 44273WS1, 44274WS9, 44275WS6, 44276WS4, 44277WS2, 44532WS0, 44533WS8, 44817WS5, 44819WS1, 44820WS9, 44821WS7, 44822WS5, 44825WS8, 44826WS6, 44908WS2, 44909WS0, 44912WS4, 44914WS0, 44915WS7, 44989WS2, 44990WS0, 44991WS8, 44992WS6, 44994WS2, 45088WS2, 45089WS0, 5122WS7
- 510(k) numbers
- ["K020263"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- not on file
- Firm city
- not on file
- Firm state
- not on file
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28