Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare Clinical Systems: GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201

Recall number
Z-1155-2008
Recalling firm
GE Healthcare Clinical Systems
Product
GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201
Status
Terminated
Initiated
2007-09-04
Posted
2008-06-12
Terminated
2012-05-29
Reason
Inaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt by the user to activate the biopsy guidelines while this probe is in use will cause incorrect guidelines to be displayed on the image. The incorrect guidelines are recognized by their orientation that is left/right reversed from t
Root cause
Software change control
Action
GE sent consignees a "GE Urgent Medical Device Correction" letter dated 8/14/07. The letter informed the customers about the problem, offer a short-term workaround and informed them that a representative will schedule a visit to upgrade the software.
Quantity
62
Distribution
Worldwide Distribution - USA (NY & AL), Korea, China, India, Brazil, Mexico, Germany, Greece, Italy, Poland, Russia, Saudi Arabia, Spain, and Czech Republic.
Lot, serial or model codes
Systems with software version 4.0.5, 4.1.1 and 4.1.2. Serial numbers: 26022WS4, 26023WS2, 26346WS7, 2641WS8, 31001WS1, 34381WS4, 36460WS4, 36461WS2, 43734WS3, 43735WS0, 43736WS8, 43737WS6, 43994WS3, 43995WS0, 43998WS4, 44000WS8, 44001WS6, 44002WS4, 44003WS2, 44004WS0, 44005WS7, 44011WS5, 44012WS3, 44013WS1, 44015WS6, 44016WS4, 44018WS0, 44019WS8, 44020WS6, 44100WS6, 44153WS5, 44157WS6, 44159WS2, 44270WS7, 44273WS1, 44274WS9, 44275WS6, 44276WS4, 44277WS2, 44532WS0, 44533WS8, 44817WS5, 44819WS1, 44820WS9, 44821WS7, 44822WS5, 44825WS8, 44826WS6, 44908WS2, 44909WS0, 44912WS4, 44914WS0, 44915WS7, 44989WS2, 44990WS0, 44991WS8, 44992WS6, 44994WS2, 45088WS2, 45089WS0, 5122WS7
510(k) numbers
["K020263"]
PMA numbers
not on file
Product code
IYN
Firm FEI
not on file
Firm city
not on file
Firm state
not on file
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register