Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.

Recall number
Z-1153-2024
Recalling firm
Boston Scientific Corporation
Product
EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.
Status
Open, Classified
Initiated
2023-12-19
Posted
2024-02-16
Terminated
not on file
Reason
Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.
Root cause
Under Investigation by firm
Action
On December 19, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE PRODUCT REMOVAL letters. Customers were instructed to cease further distribution or use of any remaining affected product. Distributors should forward the notice to their customers to ensure the recall is carried out to the end-user level. Customers were instructed to discontinue use and segregate affected product. Customers will receive a Return Goods Authorization (RGA) number after completing and returning their Reply Verification Tracking Form (RVTF). Customers will receive credit for returned affected product. Reach out to your local Boston Scientific representative with any questions.
Quantity
0 US; 74 OUS
Distribution
International distribution to the country of Japan only.
Lot, serial or model codes
UPN M00509500, GTIN (UDI-DI) 08714729902713, Lot 31073152; UPN M00509510, GTIN (UDI-DI) 08714729902720, Lots 31221751, 31761882, 32409573.
510(k) numbers
K161003
PMA numbers
not on file
Product code
PIF
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register