Respironics California, LLC: Respironics RP-UI Assy, V60 Ventilators User Interface RP-Kits, REF/P/N: 453561528931, 453561528941, 453561536451, 453561536461

Recall number
Z-1153-2020
Recalling firm
Respironics California, LLC
Product
Respironics RP-UI Assy, V60 Ventilators User Interface RP-Kits, REF/P/N: 453561528931, 453561528941, 453561536451, 453561536461
Status
Open, Classified
Initiated
2019-09-11
Posted
not on file
Terminated
not on file
Reason
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.
Root cause
Other
Action
On 09/26/19, Urgent Medical Device Correction Field Safety Notices were mailed to customers. Firm is asking customers to follow the Action to be Taken by Customer/User section of the Medical Device Correction notification/Field Safety Notice (Ventilator): 1) If in use on a patient, continue to use the ventilator if an alternative non-invasive ventilator is not available. 2) If a touchscreen failure is detected, disconnect the patient and immediately start ventilation with an alternate device. Contact your local customer service representative to report the failure. 3) To minimize the risk of patient injury, operate the Ventilator as defined in the operator's manual and promptly attend to all alarms presented by the ventilator. 4) As recommended in the operator's manual, use an external 02 monitor/analyzer and set the alarm thresholds appropriately. 5) As recommended in the operator's manual, have an alternative means of ventilation available whenever the ventilator is in use. 6) Acknowledge receipt of this notification. Firm is asking customers to follow the Action to be Taken by Customer/User section of the Medical Device Correction notification/Field Safety Notice (User Interface RP-Kits): 1) If the affected kit is installed on a ventilator and in use on a patient, continue to use the ventilator if an alternative non-invasive ventilator is not available. 2) If the affected kit is installed on a ventilator, document the serial number of the ventilator and the serial number of the affected User Interface Assembly it contained, on the attached Acknowledgement and Receipt Form. 3) If the affected devices could not be found or linked to a ventilator at your site, please indicate this within in the detail section provided on the form, 4) Return the completed and signed Acknowledgement and Receipt Form. Firm is asking customers to follow the Action to be Taken by Customer/User section of the Medical Device Correction notification/Field Safety Notice (Tou
Quantity
56
Distribution
US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam
Lot, serial or model codes
Serial Numbers: 100270400, 100270407, 100256220, 100270404, 100263004, 100266507, 100270402, 100263001, 100263009, 100258193, 100258197, 100258883, 100262978, 100263006, 100258196, 100256196, 100256197, 100256198, 100263003, 100266499, 100266505, 100258889, 100256192, 100258887, 100266509, 100270408, 100262976, 100258881, 100258880, 100266511, 100270406, 100258885, 100256223, 100263000, 100258884, 100258888, 100266500, 100266504, 100258886, 100256222, 100256195, 100256199, 100266502, 100266508, 100266510, 100270403, 100270490, 100257708, 100257713, 100257716, 100256831, 100256834, 100256837, 100262373, 100262556, 100262575, 100265136
510(k) numbers
["K082660"]
PMA numbers
not on file
Product code
MNT
Firm FEI
2518422
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register