Beckman Coulter Inc.: AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
- Recall number
- Z-1151-2018
- Recalling firm
- Beckman Coulter Inc.
- Product
- AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
- Status
- Terminated
- Initiated
- 2018-01-18
- Posted
- not on file
- Terminated
- 2023-09-27
- Reason
- A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.
- Root cause
- Device Design
- Action
- A recall notification letter was sent to affected US consignees on 1/19/2018 notifying them of the issue. In addition, the recalling firm is instructing field service engineers to inspect the tip present sensor for misalignment at the next service visit.
- Quantity
- 73 units in US
- Distribution
- Worldwide distribution. US nationwide, Algeria, Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
- Lot, serial or model codes
- All units in the AutoMate 2500 Family are affected; Serial Numbers - 1107, 4048, 4055, 4025, 4052, 4079, 4091, 4076, 4072, 4073, 4074, 4111, 4097, 1054, 1055, 1048, 1049, 1147, ODL25125-1342, ODL25125-1354, 1251, ODL25255-4530, ODL25255-4531, 4126, 4127, 4169, 4171, 4341, 4324, ODL25255-4536, ODL25255-4535, ODL25255-4397, 4141, 4143, 4144, 4146, ODL25255-4606, 4148, ODL25255-4577, ODL25255-4605, 4302, 4304, 4352, ODL25255-4352, 4330, ODL25255-4419, ODL25255-4534, ODL25255-4416, ODL25255-4401, 4168, 4107, 4255, 4257, 4252, 4256, 4254, 4253, 4236, ODL25255-4437, ODL25255-4443, ODL25255-4444, ODL25255-4438, 4294, 4293, ODL25255-4556, ODL25255-4382, ODL25255-4378, ODL25255-4377, ODL25255-4379, ODL25255-4380, ODL25255-4381, ODL25255-4420, ODL25255-4421
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28