FDA Medical Device Recalls (openFDA)

Megadyne Medical Products, Inc.: Hand Control 12FR Suction Coagulator with 10 ft cord and holster, Sterile, Disposable, Model Number 0043-25, packaged Tyvek polyester peel pouch, 25 pouches per box. This device is intended to conduct monopolar electrosurgical energy from an electrosurgical generator (ESU) to target tissue during ENT and general surgical procedures. It is also a suction apparatus used to remove surgical smoke and fluids from the site. The device is intended for single use; it is not intended to be cleaned or reused.

Recall number
Z-1151-2014
Recalling firm
Megadyne Medical Products, Inc.
Product
Hand Control 12FR Suction Coagulator with 10 ft cord and holster, Sterile, Disposable, Model Number 0043-25, packaged Tyvek polyester peel pouch, 25 pouches per box. This device is intended to conduct monopolar electrosurgical energy from an electrosurgical generator (ESU) to target tissue during ENT and general surgical procedures. It is also a suction apparatus used to remove surgical smoke and fluids from the site. The device is intended for single use; it is not intended to be cleaned or reused.
Status
Terminated
Initiated
2014-02-27
Posted
2014-03-07
Terminated
2014-07-09
Reason
Megadyne Medical Products, Inc. is voluntarily recalling Hand Control 12Fr Suction Coagulator with 10ft cord and 132405 Holster, Sterile, Disposable, Lot 132405 due to the potential for a small number of these suction coagulators to self-activate immediately on connection to the electrosurgical generator and to remain activated until it is disconnected.
Root cause
Use error
Action
Megadyne Medical Product sent an Urgent Medical Device Recall letter dated February 27, 2014, to all affected customers with follow-up phone calls and emails. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed toexamine their inventory immediately to determine if they have any of the affected product. If so, customers were instructed to remove the product from service and contact Megadyne Customer Service at 1-801-576-9669 or 1-800-747-6110 for return authorization and replacement or credit. If customers have distributed any the affected product, they were instructed to immediately contact their distribution points to have the product returned to Megadyne. Customers with questions were instructed to contact Customer Service at 1-800-747-6110 or 1-801-576-9669 or email Regulatory @Megadyne.com. For questions regarding this recall call 801-576-9669.
Quantity
494
Distribution
USA including FL, MN, LA, NJ, NM, PA, TX and Internationally to Canada.
Lot, serial or model codes
Catalog Number: 0043-25 Lot Number: 132405 Exp. Date: 07-2018
510(k) numbers
["K072559"]
PMA numbers
not on file
Product code
GEI
Firm FEI
1721194
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register