Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Infusion Pump

Recall number
Z-1150-2021
Recalling firm
Baxter Healthcare Corporation
Product
Infusion Pump
Status
Terminated
Initiated
2021-01-06
Posted
not on file
Terminated
2026-05-04
Reason
Battery performance issues related to prolonged storage.
Root cause
Under Investigation by firm
Action
1. If your facility is currently installing or planning to install new pumps and batteries that are currently in storage, please contact Baxter Technical Assistance at 800 356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time, for further instructions. The duration and conditions under which the new batteries have been stored will help determine whether battery damage due to depletion is likely. 2. For facilities currently using Spectrum pumps: If you are not experiencing recurrent battery alerts/alarms, continue using the Spectrum battery modules as indicated. If you are experiencing recurrent battery alerts/alarms, please operate your pumps on AC power and contact Baxter Technical Assistance at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time. 3.Baxter is advising all customers who plan prolonged storage of Spectrum pumps and batteries to ensure the following: Batteries are charged to 2 or 3 bars prior to storage (do not fully charge); Batteries are stored between -10 and +35 degrees C (14 and 95 degrees F); Batteries are removed from the pump and stored independent of the pumps; and Batteries are recharged every six months while in storage. Not following these storage instructions can result in battery damage and/or reduced capacity and life of the battery. 4. The Spectrum Installation & Maintenance Guide/Service Manual, which can be accessed at https://service.baxter.com/tsportal/, describes a Battery Life Test that can be used to test for battery damage due to depletion. This test is recommended as part of routine maintenance. 5. Customers experiencing recurrent battery alerts/alarms should follow the instructions described in step 2 above. Do not attempt to resolve this issue by performing the battery recovery as described in the Technical Service Bulletins below. The battery recovery process is intended to recover a battery locked in a low c
Quantity
589,704 total units
Distribution
Domestic Distribution: ST, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Puerto Rico. International Distribution: Canada, Barbados, Bermuda, St. Thomas, Guyana, Trinidad and Tobago, Jamaica and Nassau Bahamas.
Lot, serial or model codes
Infusion Pump Product codes: 35700ABB, 35700BAX, and 35700BAX2  Battery codes: 35083, 35162, 35195, and 35724.  Infusion Pump Product codes: 35700ABB, 35700BAX, 35700BAX2 3570009  Battery codes: 35223 and 36010
510(k) numbers
["K042121","K133801","K173084"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1314492
Firm city
Medina
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register