Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: AXIOM Artis angiographic x-ray system, Model Numbers: 7412807, 7413078, 7728392 AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Recall number
Z-1150-2018
Recalling firm
Siemens Medical Solutions USA, Inc
Product
AXIOM Artis angiographic x-ray system, Model Numbers: 7412807, 7413078, 7728392 AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Status
Terminated
Initiated
2018-01-05
Posted
2018-03-20
Terminated
2018-08-23
Reason
Certain systems may have leaking of the plastic coolant couplings, the liquid can enter into the flat panel detector leading to a failure of the detector. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
Root cause
Process control
Action
The firm, Siemens Healthineers, sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated 1/5/2018 to its consignees. The letter describes the product, problem and actions to be taken. The consignees were instructed to follow their standard emergency procedures in place in the event of system failures; to consider reinforcing these emergency procedures until the update has been completed. Siemens will replace all affected plastic couplings with metal couplings. The Siemens service organization will contact the consignees to arrange a date to perform this corrective action.If the device has been sold and is therefore no longer in your possession, please forward this field correction notice to the new owner. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
Quantity
22
Distribution
US Distribution in the following U.S. states: AZ, CA, FL, IA, IL, LA, MI, NC, NJ, OH, OK, and TX.
Lot, serial or model codes
Serial Numbers: 35061 35054 34999 35098 35059 35064 35073 35071 35057 35075 35083 35018 35097 35108 35046 35048 35050 46004 35045 35004 44011 35095
510(k) numbers
["K021021"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register