Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: OT1000 Series Orthopedic Surgical Tables

Recall number
Z-1149-2021
Recalling firm
Steris Corporation
Product
OT1000 Series Orthopedic Surgical Tables
Status
Terminated
Initiated
2021-01-21
Posted
not on file
Terminated
2021-11-10
Reason
On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Root cause
Process control
Action
In January, 2021, Steris will issue an Urgent Medical Device Correction notice to customers via letter for the OT1000 Series Orthopedic Surgical Table. Steris has identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Quantity
272 tables total (Domestic: 227 tables, Foreign 45 tables)
Distribution
Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.
Lot, serial or model codes
Model No.: OT1000, OT1100, OT1200
510(k) numbers
not on file
PMA numbers
not on file
Product code
JEA
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.