FDA Medical Device Recalls (openFDA)
Medefil Incorporated: Heparin I.V. Flush Syringe is available is pre-filled polypropylene syringes, 100 units/mL, 3 mL fill in a 12 mL syringe.
- Recall number
- Z-1149-2014
- Recalling firm
- Medefil Incorporated
- Product
- Heparin I.V. Flush Syringe is available is pre-filled polypropylene syringes, 100 units/mL, 3 mL fill in a 12 mL syringe.
- Status
- Terminated
- Initiated
- 2014-02-19
- Posted
- 2014-03-06
- Terminated
- 2017-09-05
- Reason
- The bar code that is printed on the Master Carton label does not correspond to the product code fully displayed on the Master Carton label. The non-matching information can only be detected by using a bar code reader. The defective bar codes will be read as Product No. MIH-3335 ( Heparin I.V. Flush Syringe 100 units/mL; 5 mL fill in 12 mL Syringe) whereas this product is actually Product No. MIH-3333 (Heparin I.V. Flush Syringe 100 units/mL; 3 mL fill in 12 mL Syringe).
- Root cause
- Labeling Change Control
- Action
- URGENT MEDICAL DEVICE CORRECTION letters dated February 19, 2014 were sent to all direct accounts. The letters included instructions to: quarantine the recalled products; and,complete and return the attached customer response form (via Fax to 630-681-9100 or via mail to Medefil Inc., 250 Windy Point Drive, Glendale Heights, IL 60139) so that the correct number of replacement labels can be sent from Medefil.
- Quantity
- 1,608 Master Cartons
- Distribution
- USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, SC, TX, UT, and WI, and the country of Canada.
- Lot, serial or model codes
- Product Number: MIH-3333; NDC Number: 64253-333-33; Lot Numbers and Expiration Dates: Lot H114120N, Expires 12/15; Lot H113403N, Expires 11/15; Lot H113369N, Expires 10/15; Lot H113327N, Expires 8/15; Lot H113291N, Expires 7/15; Lot H113266N, Expires 6/15; Lot H113232N, Expires 5/15; Lot H113185N, Expires 3/15.
- 510(k) numbers
- ["K092491"]
- PMA numbers
- not on file
- Product code
- NZW
- Firm FEI
- 3001677091
- Firm city
- Glendale Heights
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28