FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Arts one include cardiac angiography, neuro-angiography, general angiography, rotational angiography; multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
- Recall number
- Z-1148-2018
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Arts one include cardiac angiography, neuro-angiography, general angiography, rotational angiography; multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
- Status
- Terminated
- Initiated
- 2018-01-12
- Posted
- 2018-03-20
- Terminated
- 2018-08-28
- Reason
- Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.
- Root cause
- Process control
- Action
- Siemens sent an Urgent - Medical Device Correction letter dated January 5, 2018. Siemens will conduct an inspection of all affected Artis tables and verify the correct positioning of the bolt as well as the snap rings. The letter recommends the consignees follow Please follow the standard emergency procedures they have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. The Siemens service organization will contact the consignees to arrange a date to perform the corrective action. For further questions, please call (610) 448-6461.
- Quantity
- 16
- Distribution
- US Distribution to the states of : AL, AR, CA, FL, IN, LA, MI, MO, NC, NY, PA, TX, and WA.
- Lot, serial or model codes
- Serial numbers: 82044 82105 82070 82213 82054 82040 82038 82205 82065 82267 82108 82010 82039 82109 82102 82066
- 510(k) numbers
- ["K133580"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28