Skip to the content

FDA Medical Device Recalls (openFDA)

Northwest Medical Physics Equipment Inc: ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software.

Recall number
Z-1147-2007
Recalling firm
Northwest Medical Physics Equipment Inc
Product
ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software.
Status
Terminated
Initiated
2005-05-31
Posted
2007-08-08
Terminated
2011-04-19
Reason
When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
Root cause
Software design
Action
Software to fix the described problems was written and customers contacted by e-mail on 05/31/2005, that a software patch, titled 05312005 was available on the firm's internet website.
Quantity
85
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Lot, serial or model codes
ISOLOC Version 4.2 Part Number NT-NW-425-520.
510(k) numbers
["K904908"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1000126033
Firm city
Everett
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register