Skip to the content

FDA Medical Device Recalls (openFDA)

Vascular Insights, LLC: ClariVein IC infusion catheter, Model Number: 65-018-E4S Product Usage: The ClariVein IC infusion catheter is indicated for infusion of physician-specified agents into the peripheral vasculature.

Recall number
Z-1146-2018
Recalling firm
Vascular Insights, LLC
Product
ClariVein IC infusion catheter, Model Number: 65-018-E4S Product Usage: The ClariVein IC infusion catheter is indicated for infusion of physician-specified agents into the peripheral vasculature.
Status
Terminated
Initiated
2017-11-09
Posted
not on file
Terminated
2018-12-18
Reason
Certain catheters could have a compromised catheter shaft due to a manufacturing defect which could result in some delamination of the catheter shaft including during introduction of the catheter into a vascular access introducer.
Root cause
Process control
Action
Vascular Insights sent an Urgent Voluntary Medical Device Recall letter to their customers. The letter identified the affected product, problem and the actions to be taken. Customers were instructed to 1. Do not use or distribute any recalled product. 2. Immediately remove all recalled product from your inventory. 3. Segregate recalled product in a secure location and follow the instruction herein to return the recalled product and receive replacement product. 4. Immediately forward a copy of this recall notification to any/all sites to which you may have distributed the affected product and provide directions to follow the instructions contained herein. Customers were asked to complete and return the enclosed Reply Verification Tracking Form.
Quantity
928
Distribution
U.S Nationwide in the states: AZ, CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MO, NE, NJ, NY, OH, PA, SC, TN, TX, VA, WA
Lot, serial or model codes
Lot number: BM651403017I
510(k) numbers
["K071468"]
PMA numbers
not on file
Product code
KRA
Firm FEI
3005831739
Firm city
Quincy
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register