Skip to the content

FDA Medical Device Recalls (openFDA)

Ventana Medical Systems Inc: VENTANA PATHWAY anti-HER-2/neu (4B5) RABBIT MONOCLONAL PRIMARY ANTIBODY (PATHWAY HER2 (4B5)); Model Number: 790-2991, Lot number: 680957, in vitro diagnostic.

Recall number
Z-1146-2010
Recalling firm
Ventana Medical Systems Inc
Product
VENTANA PATHWAY anti-HER-2/neu (4B5) RABBIT MONOCLONAL PRIMARY ANTIBODY (PATHWAY HER2 (4B5)); Model Number: 790-2991, Lot number: 680957, in vitro diagnostic.
Status
Terminated
Initiated
2009-09-17
Posted
2010-03-16
Terminated
2010-04-22
Reason
Incorrect result: Light staining inconsistencies with lot number 680957 of the Ventana PATHWAY HER-2/neu (4B5) Rabbit Monoclonal Primary Antibody.
Root cause
Nonconforming Material/Component
Action
The recall was initiated on 09/17/2009 with Ventana Medical Systems forwarding Urgent Voluntary Medical Device Recall letters to all the affected customers advising them to discontinue the use of lot number 680957 and have their laboratory retest with product from a different lot of PATHWAY HER2 (4B5) any breast cases scored as 0, 1+, or 2+ that were not reflexed to an approved in situ hybridization (ISH) method. Customers were also instructed to contact Ventana Customer Support at 1-800-227-2155 and select 1 for Technical Support to speak to a Customer Support agent who will work with then to replace or credit any product from this specific lot number.
Quantity
433 units
Distribution
Worldwide distribution: USA, Austria, Denmark, Germany, Hungary, Netherlands, Sweden, United Kingdom, Brazil, Canada, India, Korea, Morocco & Thailand
Lot, serial or model codes
Lot number: 680957
510(k) numbers
not on file
PMA numbers
["P990081"]
Product code
MVC
Firm FEI
1000125183
Firm city
Tucson
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register